Module 07 · ABR domain III.A · Treatment planning

Photon Dosimetry and MU Calculations

Every plan ends in a number of monitor units. This module explains what that number means, every factor that goes into it, and how to check it by hand in under two minutes. It is the single most calculation-heavy topic on the exam and the one physicists assume you already know.

In this module
  1. 01The one idea
  2. 02Monitor units and calibration
  3. 03Inverse square
  4. 04Buildup and dmax
  5. 05Percent depth dose
  6. 06TAR, TPR, TMR
  7. 07Field size factors
  8. 08Beam modifiers
  9. 09The two equations
  10. 10MU calculator
  11. 11Worked examples
  12. 12In the planning room
  13. 13Board pearls and traps
  14. 14Question bank
  15. 15Formula card and sources
01 · Start here

The one idea

A monitor unit calculation answers one question: how much dose does one MU deliver to my point, in my patient, with my setup? Divide the prescribed dose by that number and you have the MU.

MU = Prescribed dose at the pointDose per MU at that point
The machine is calibrated to give a known dose per MU under one reference condition. Every factor in a hand calc converts that reference dose per MU to your actual condition.

Think of calibration as a single known exchange rate: 1 MU buys 1 cGy, in water, at a fixed distance, at the depth of maximum dose, in a 10 x 10 cm field. Your patient differs from that reference in four ways, and each difference gets one correction factor:

  1. Distance from the source to the point. Corrected by the inverse square factor.
  2. Depth of tissue the beam crosses before the point. Corrected by PDD or TMR.
  3. Field size, which changes how much scattered radiation reaches the point. Corrected by Sc and Sp.
  4. Things in the beam such as wedges and block trays. Corrected by their transmission factors.

Every one of these factors sits in the denominator. If a factor is below 1 (deeper point, smaller field, a wedge), dose per MU falls, so you need more MU. If you remember nothing else, remember that direction. It lets you catch most calculation errors without finishing the arithmetic.

Pearl

Before you calculate, predict. "Deep point, small field, wedge in: MU will be well above the dose in cGy." A prediction you make first protects you from a decimal slip.

02 · The unit itself

Monitor units and calibration

An MU is not a dose. It is a count from the ionization chambers in the treatment head, called monitor chambers, which sit in the beam path below the flattening filter (or where it would be, in flattening-filter-free beams). The beam passes through them, they collect charge, and the machine stops when the count reaches the set MU. Physics then tunes the chamber electronics so that, under reference conditions, 1 MU produces 1 cGy.

There are two monitor chambers, a primary and a secondary. If the primary fails, the secondary terminates the beam a few MU later. A backup timer stands behind both. Modern chambers are sealed, so their reading does not drift with room temperature and pressure.

How the machine gets calibrated

Absolute calibration follows the AAPM TG-51 protocol. Physics places a calibrated ion chamber in a water tank at 10 cm depth, 10 x 10 cm field, at either 100 cm SSD or 100 cm SAD, measures charge, and converts it to dose to water using the chamber's calibration coefficient and a beam quality factor, kQ. Beam quality for photons is specified by %dd(10)x, the photon-only percent depth dose at 10 cm. The clinic then uses its own depth dose data to carry that measured dose back up to dmax, where the "1 cGy per MU" convention is usually defined.

Board point

TG-51 reference depth is 10 cm, not dmax, because at 10 cm the dose is free of the electron contamination that distorts the buildup region. Beam quality is %dd(10)x, measured at 100 cm SSD. For beams of 10 MV and above, a thin lead foil is placed in the beam during that measurement to account for contaminant electrons.

The convention that changes your arithmetic

Clinics define their reference point in one of two ways. You need to know which, because it decides the inverse square factor in an isocentric calculation.

ConventionReference pointSource to calibration pointISF for an isocentric calc (6 MV)
SSD-calibrateddmax, with the water surface at 100 cm100 + dmax = 101.5 cm(101.5 / 100)² = 1.030
SAD-calibrateddmax, with that point at isocenter100 cm(100 / 100)² = 1.000
In clinic

Ask your physicist once: "Are our machines calibrated SSD or SAD, and at what depth?" Write it on your formula card. The answer changes every isocentric hand calc by about 3 percent at 6 MV and more at higher energy, because dmax is deeper.

03 · Distance

Inverse square

The photon beam spreads from a nearly point-like source. The same number of photons passes through every cross-section of the beam, but the cross-section's area grows with the square of the distance. Photon fluence, and the dose it produces in air, therefore falls with the square of distance from the source.

D2 = D1 × ( r1r2 )²
r is measured from the source, not from the skin. Dose at the farther point is smaller, so the ratio is always the near distance over the far distance when you move away.

Example: a point at 100 cm receives 100 cGy. At 110 cm it receives 100 × (100/110)² = 82.6 cGy. Ten percent farther away costs about 17 percent of the dose.

Inverse square describes the primary beam in air. Inside a patient, attenuation and scatter also act, which is why we need depth dose tables. Inverse square still appears inside every MU equation, but only to account for the change in distance between the calibration point and your point.

Pearl

Near 100 cm, a 1 cm change in distance changes dose by about 2 percent: (100/101)² = 0.980. A patient set up 1.5 cm closer to the source than planned on an SSD treatment gets about 3 percent more dose at depth. This is the number behind every SSD check on day one.

04 · The first centimetres

Buildup and dmax

Photons do not deposit dose directly. They hand their energy to electrons (mostly by Compton scattering at therapy energies), and those electrons deposit dose along their tracks as they slow down. Megavoltage photons throw electrons mainly forward, and those electrons travel centimetres before stopping.

At the skin, few electrons have been set in motion yet, so dose is low. With each millimetre, electrons generated above add to those generated here. Dose rises until the number of electrons entering a small volume equals the number leaving it. That balance is called charged particle equilibrium, and the depth where dose peaks is dmax. Past dmax, the photon beam is being attenuated, fewer electrons are produced, and dose falls.

Because higher-energy electrons travel farther, dmax moves deeper as beam energy rises. That is the physical basis of skin sparing, and the reason high-energy beams spare skin better.

Beamdmax (approx.)PDD at 10 cm, 10 x 10Clinical use
Co-600.5 cm≈ 56% (at 80 cm SSD)Historical; Gamma Knife source
4 MV1.0 cm≈ 62%Older units, thin sites
6 MV1.5 cm≈ 67%Workhorse: HN, breast, extremity, most VMAT
10 MV2.3 to 2.5 cm≈ 73%Pelvis, thicker patients
15 MV≈ 2.9 cm≈ 77%Large separations
18 MV3.0 to 3.5 cm≈ 79%Large separations; produces neutrons

Values are typical for 100 cm SSD unless noted. Your machine's commissioned data take precedence; these are the numbers to know for the exam.

What raises surface dose

Anything that adds electrons above the skin: larger fields (more electrons scattered from the head and air), a block tray or anything else close to the skin, bolus, and oblique incidence. Surface dose for a 6 MV 10 x 10 field at normal incidence is roughly 15 to 25 percent; it is lower at higher energy.

Board point

dmax shifts slightly shallower as field size increases, because larger fields carry more contaminating electrons from the treatment head. Bolus moves dmax up toward the skin by its thickness. Placing a tray close to the skin destroys skin sparing, which is why trays sit at least about 15 to 20 cm above the skin.

In clinic

When you want full dose on a mastectomy scar or a superficial node, you are asking to defeat buildup. Bolus is the tool. When you want to protect skin in a deep pelvic target, buildup is your friend and higher energy helps.

05 · Fixed SSD geometry

Percent depth dose

Percent depth dose (PDD) is the dose at depth d on the central axis divided by the dose at dmax on the central axis, times 100, for one beam at a fixed source-to-surface distance (SSD). The point and the reference point are at different distances from the source. That detail is why PDD depends on SSD and why it is the natural quantity for SSD setups.

PDD(d, r, SSD, E) = 100 × D at depth dD at dmax
r is the field size defined at the surface.
Increase thisPDD beyond dmaxWhy
DepthFallsAttenuation and inverse square both act
EnergyRisesHigher-energy photons are less attenuated in water
Field sizeRisesMore scatter reaches the deep point, relatively more than at dmax
SSDRisesThe fractional distance between dmax and the deep point shrinks, so inverse square costs less

The SSD effect is the one residents miss. At 100 cm SSD the dmax point is at 101.5 cm and a 10 cm point at 110 cm, an inverse square ratio of (101.5/110)² = 0.851. At 120 cm SSD the same depths sit at 121.5 and 130 cm, a ratio of 0.874. The deep point loses less to distance, so its PDD is higher. The absolute dose rate is lower at extended SSD, but the shape of the depth dose curve is more penetrating.

Depth dose explorer

Teaching model, not commissioning data
Energy

Dashed line: the same energy at 10 x 10, 100 cm SSD, for reference. Move one slider at a time and say out loud why the curve moved.

PDD, your settingsPDD, 10 x 10 at 100 cm SSD
dmax1.5 cm
Surface0%
PDD at 5 cm0%
PDD at 10 cm0%
PDD at 20 cm0%
TMR at 10 cm0

Correcting PDD for a different SSD: the Mayneord F factor

PDD tables are measured at one SSD, usually 100 cm. For an extended SSD treatment (large fields, TBI, a patient who does not fit), you scale the table value by the Mayneord F factor, which corrects only the inverse square part of the change.

F = ( SSD2 + dmaxSSD1 + dmax )² × ( SSD1 + dSSD2 + d )²
PDD(SSD2) ≈ PDD(SSD1) × F
For 6 MV at 10 cm depth, moving from 100 to 120 cm SSD: F = (121.5/101.5)² × (110/130)² = 1.026, so 67% becomes about 68.7%.
Board point

The F factor ignores the change in scatter. It works well for small fields and slightly overstates the PDD increase for large fields and deep points, where scatter makes up more of the dose. F is greater than 1 when SSD increases.

06 · Fixed point geometry

TAR, TPR and TMR

Modern treatments are isocentric. The machine rotates around a fixed point, the isocenter, 100 cm from the source, and we place the tumor there. From one beam to the next the point stays at 100 cm; what changes is how much tissue lies above it. PDD is clumsy here because it ties depth and distance together. We need a ratio where the point stays at the same distance from the source and only the depth of overlying tissue changes. Three such ratios exist.

SSD setup: surface fixed at 100 cm SAD setup: point fixed at 100 cm source water / patient point at depth d SSD = 100 d point sits at 100 + d from source source water / patient isocenter, at depth d SAD = 100 d surface sits at 100 − d (the SSD you read on the ODI)

In SSD geometry the point moves away from the source as depth increases, so PDD carries an inverse square fall-off. In SAD geometry the point never moves, so TMR carries attenuation and scatter only.

RatioDefinition (same point, same distance from source)Where you meet it
TARDose at depth d in phantom ÷ dose at the same point in free space (in air, with just enough buildup)Co-60 era; awkward at MV because "in air" measurement needs a large buildup cap
BSF / PSFTAR at dmax; how much phantom scatter adds at the peak. Called peak scatter factor at MVKilovoltage and orthovoltage, where it can exceed 1.3
TPRDose at depth d ÷ dose at a reference depth, same pointGeneral form; reference depth often 10 cm
TMRTPR with the reference depth at dmaxStandard for isocentric hand calcs and many planning systems
SARTAR(d, r) − TAR(d, 0): the scatter part onlyClarkson integration for irregular fields

TMR depends on energy, depth and field size at the point. It does not depend on SSD, because both doses in the ratio are measured at the same distance from the source, so inverse square cancels. That independence is exactly why it works for isocentric setups.

TMR is always at least as large as PDD at the same depth and field size beyond dmax, because PDD contains an extra inverse square fall-off. At 6 MV, 10 x 10, 10 cm deep: TMR ≈ 0.79 while PDD ≈ 67%.

PDD(d, r, f) = 100 × TMR(d, rd) × Sp(rd)Sp(rdmax) × ( f + dmaxf + d )²
Converting between the two. rd and rdmax are the field size projected to depth d and to dmax. You will rarely compute this by hand, but its parts explain every difference between PDD and TMR: an inverse square term and a small scatter term.
Board point

Quantities independent of SSD: TAR, TPR, TMR, SAR. Quantity that depends on SSD: PDD. TMR at dmax equals 1 by definition. In Clarkson's method, the primary (zero-area TAR) and the scatter (SAR summed over radial sectors) are calculated separately and added.

07 · Where most errors live

Field size factors

Open the jaws and two things increase. More of the treatment head (mainly the flattening filter) is visible from the point, so more photons scatter out of the head toward it. And more patient tissue is irradiated, so more photons scatter within the patient toward it. The two are separated because blocking and MLC shaping change them differently.

FactorWhat it capturesWhich field size to useHow it is measured
Sc collimator scatter, head scatterScatter from the head; output in airCollimator (jaw) setting, defined at isocenter. Blocks and MLC do not change it (to first order)In air, chamber with buildup cap or miniphantom, normalized to 10 x 10
Sp phantom scatterScatter from the irradiated patient or waterActual irradiated (blocked) field at the point of calculationDerived: Sp = Sc,p ÷ Sc
Sc,p total scatter factor, output factorBoth togetherOpen field with no blockingIn water at reference depth, normalized to 10 x 10
Trap

A 20 x 20 jaw opening with MLC shaping down to an equivalent 12 x 12 field: Sc uses 20 x 20, Sp uses 12 x 12. Using 12 for both underestimates the head scatter and overstates the MU. Using 20 for both overstates the patient scatter and under-doses. Exams test this directly.

Rectangles and circles: equivalent square

Tables are indexed by square fields. A rectangular field gives the same central axis scatter as a square field of the same area-to-perimeter ratio (Sterling's rule).

sideeq = 4 × AreaPerimeter = 2aba + b
Circle of diameter D: sideeq ≈ 0.89 × D (equal area)
6 x 20 field: 2 × 6 × 20 / 26 = 9.2 cm. The equivalent square is always closer to the short side than you expect, because the short side dominates scatter.

Typical sizes of the effect

For a flattened 6 MV beam, Sc runs roughly from 0.96 at 4 x 4 to about 1.05 at 40 x 40; Sp spans a similar range. A field-size error of a few centimetres therefore moves dose by 1 to 3 percent. Small fields (below about 4 x 4) are a different regime: the source is partly hidden, lateral equilibrium is lost, and output drops steeply. That belongs to the SRS module.

Board point

Collimator exchange effect: a 5 x 20 field and a 20 x 5 field give slightly different Sc. The upper jaw, closer to the source, has more influence on head scatter, so Sc is larger when the upper jaw defines the long side. Sp is identical for both.

Pearl

Flattening-filter-free beams have much less head scatter, so their Sc changes less with field size. They also deliver higher dose rates (on a TrueBeam, up to 1400 MU/min at 6 MV FFF and 2400 MU/min at 10 MV FFF, versus 600 MU/min flattened).

08 · Things in the beam

Beam modifiers

Anything placed in the beam is handled with a transmission factor: dose on the central axis with the device divided by dose without it, same MU. It sits in the denominator with everything else.

DeviceTypical factorWhat else it changes
Block tray≈ 0.95 to 0.98Adds surface dose if close to skin
Physical wedge≈ 0.5 to 0.8, angle dependentHardens the beam, so PDD rises slightly; factor varies with depth and field size
Enhanced or dynamic wedgevaries with field lengthA jaw sweeps during delivery. No hardening; factor depends strongly on field size along the sweep direction
MLCbuilt into SpLeakage and transmission of 1 to 2 percent through closed leaves; matters for heavily modulated plans
Board point

The wedge angle is the angle by which an isodose line (classically the 50 percent line, or the line at 10 cm depth by ICRU) is tilted from perpendicular to the central axis. Hinge angle θ between two beams calls for a wedge angle of 90° − θ/2. Two beams at 90° need 45° wedges.

09 · Put it together

The two equations

Every hand calc on the exam is one of these two equations. Read them as the "one idea" from section 01 with every correction written out.

Non-isocentric, fixed SSD
MU = DRxK × PDD(d, rs, SSD)100 × Sc(rc) × Sp(rdmax) × WF × TF × ISFSSD
ISFSSD = ( SCD ÷ (SSD + dmax) )²
Equals 1 when the machine is SSD-calibrated and you treat at 100 cm SSD. For extended SSD, correct the PDD with the F factor as well.
Isocentric, fixed SAD
MU = DRxK × TMR(d, rd) × Sc(rc) × Sp(rd) × WF × TF × ISFSAD
ISFSAD = ( SCD ÷ SAD )²
SCD is 101.5 cm for a 6 MV SSD-calibrated machine and 100 cm for an SAD-calibrated one.
KCalibrated output, usually 1 cGy/MU at the reference point
SCDSource to calibration point distance
rcCollimator (jaw) setting at isocenter, as an equivalent square
rsBlocked field size at the surface (SSD calc)
rdBlocked field size at the point of calculation (SAD calc; at isocenter)
rdmaxBlocked field size at dmax; for SSD calcs at 100 cm, effectively the surface field
WF, TFWedge factor and tray factor; 1 if absent
Pearl

To find dose from MU, flip the equation: D = MU × K × (all the factors). Exams love "a field delivered X MU, what dose did point Y receive?" questions. The factors are identical; only the algebra moves.

10 · Run it

MU calculator

This runs the equations above on a teaching model of a 6 MV and an 18 MV beam. It shows every factor, so you can change one input and watch which factor moves. Use it to check your own hand calcs from the worked examples; never use it for a patient.

Hand calc bench

Teaching model, not for clinical use
Setup

Jaw settings are collimator settings at 100 cm, as on the console.

0MU
11 · Practice the method

Worked examples

Work each one on paper before opening the solution. Factors are given in the problem, as they are on the exam; the method is what you are practising.

EX 1

Single field, fixed SSD

6 MV, 10 x 10 cm at the surface, 100 cm SSD. Prescribe 200 cGy at 5 cm depth. PDD(5) = 86.5%. Machine calibrated 1 cGy/MU at dmax, 100 cm SSD. No modifiers.

Show solution

Reference field, reference SSD: Sc = Sp = 1 and ISF = 1. Only depth changes.

MU = 200 ÷ (1 × 0.865) = 231.2 MU

≈ 231 MU

Sanity check: dose at dmax is 231 cGy, which is 200 ÷ 0.865. Every SSD field delivers more dose at dmax than at the prescription depth.

EX 2

Single field, isocentric

6 MV, 10 x 10 cm at isocenter, isocenter 10 cm deep. Prescribe 200 cGy to isocenter. TMR(10, 10x10) = 0.790. Machine calibrated 1 cGy/MU at dmax with 100 cm SSD.

Show solution

The calibration point is at 101.5 cm, the isocenter at 100 cm. The isocenter is closer to the source, so it gets more dose per MU: ISF = (101.5/100)² = 1.030.

MU = 200 ÷ (0.790 × 1.030) = 245.7 MU

≈ 246 MU

Compare with an SSD setup to 10 cm depth: 200 ÷ 0.67 = 298.5 MU. The isocentric point is 10 cm closer to the source (100 versus 110 cm), so it needs about 18 percent fewer MU.

EX 3

Parallel opposed, isocentric, blocked

AP/PA, 6 MV, isocenter at midplane of a 22 cm separation. 180 cGy per fraction to isocenter, equally weighted. Jaws 15 x 12, MLC blocks the field to an effective 12 x 10. Given: Sc(13.3) = 1.012, Sp(10.9) = 1.003, TMR(11, 10.9) = 0.770. SSD-calibrated machine. Find MU per field.

Show solution

Each field delivers half: 90 cGy at 11 cm depth. Collimator equivalent square: 2 × 15 × 12 / 27 = 13.3 cm, for Sc. Blocked equivalent square: 2 × 12 × 10 / 22 = 10.9 cm, for Sp and TMR.

MU = 90 ÷ (0.770 × 1.012 × 1.003 × 1.030) = 111.8 MU

≈ 112 MU per field

Both fields get the same MU only because isocenter is at midplane. If the isocenter were off midline, the AP and PA depths differ and so do their MU.

EX 4

Extended SSD

6 MV single field at 120 cm SSD. Field is 10 x 10 at the surface. Prescribe 200 cGy at 10 cm depth. PDD(10, 10x10) at 100 cm SSD = 67.0%. Sc(8.3) = 0.992, Sp ≈ 1.00. SSD-calibrated machine.

Show solution

Step 1, collimator setting: a 10 cm field at 120 cm was set as 10 × 100/120 = 8.3 cm at isocenter, so Sc(8.3) = 0.992.

Step 2, PDD at the new SSD: F = (121.5/101.5)² × (110/130)² = 1.026, so PDD = 67.0 × 1.026 = 68.7%.

Step 3, distance: the dmax point is now at 121.5 cm versus 101.5 cm at calibration. ISF = (101.5/121.5)² = 0.698.

MU = 200 ÷ (0.687 × 0.992 × 1.00 × 0.698) = 420.5 MU

≈ 421 MU

Extended SSD costs a lot of MU because inverse square dominates, even though the PDD improves slightly. That is the trade at the heart of TBI.

EX 5

Add a wedge

Repeat EX 2 with a 45° physical wedge, wedge factor 0.72.

Show solution
MU = 200 ÷ (0.790 × 1.030 × 0.72) = 341.3 MU

≈ 341 MU

A wedge factor of 0.72 raises MU by 1/0.72 = 1.39 times. If a wedged field shows the same MU as the open field, the wedge was left out of the calculation.

EX 6

Reverse: what did the skin side get?

A single 6 MV SSD field (100 cm SSD, 10 x 10) is prescribed 200 cGy at 10 cm depth. PDD(10) = 67%. What is the dose at dmax?

Show solution
D(dmax) = 200 ÷ 0.67 = 298.5 cGy

≈ 299 cGy, about 149% of the prescription

This is why single deep fields are rarely used: the entrance region is far hotter than the target. Opposed fields, multiple fields and higher energy all exist to fix that.

12 · Clinic layer

In the planning room

You will not hand-calculate a VMAT plan, but the same physics decides whether a plan's MU look reasonable, whether a setup discrepancy matters, and what to ask when a number looks wrong.

Ballpark numbers to carry in your head

SituationRough numberWhat it tells you
6 MV, 10 x 10, 200 cGy to isocenter at 10 cm≈ 245 MUDose per MU near 0.8 cGy at typical depths
Same, 18 MV≈ 210 MUHigher energy: fewer MU for the same deep dose
Plan MU far above dose in cGy for 3D fieldsMU/cGy > 1.5Deep point, small field, wedge, or an error. Find which.
VMAT or IMRT versus 3D for the same targetoften 1.5 to 3 times the MUModulation. Very high MU means more leakage dose and more sensitivity to MLC errors
SSD off by 1 cm≈ 2% at depthDistance errors matter at the percent level
Treatment timeMU ÷ dose rate300 MU at 600 MU/min is 30 s of beam-on

The independent MU check

Every plan gets a second, independent calculation of MU, by hand or with separate software, before first treatment (AAPM TG-114 covers this). For simple geometry the two should agree within a few percent; heterogeneous anatomy, small or highly blocked fields, points far off-axis and oblique surfaces justify wider tolerances, and each clinic sets its own action level. IMRT and VMAT are verified differently, through patient-specific QA measurement or an independent dose recalculation, because a single point calculation cannot describe a modulated field.

When the planning system and the check disagree, the usual reasons are predictable: lung or bone heterogeneity (the hand calc assumes water), a point near a field edge or in a steep gradient, missing tissue or oblique incidence, a wrong field size in Sc or Sp, or a forgotten wedge or tray.

Say this to physics or dosimetryWhy it lands
"What's the second-check agreement on this plan, and where's the calc point?"A good agreement at a bad point (in a gradient or in lung) means little
"Is the calc point in lung? How much of the difference is heterogeneity correction?"Shows you know a water-based check will disagree in lung, and by roughly how much
"The MU look high for this depth. Is there a wedge, or is the field small?"Uses the direction-of-factors logic from section 01
"How many MU per Gy on this VMAT? Can we lower modulation without losing coverage?"Lower MU means less leakage and a plan more robust to small MLC errors
"The patient lost weight; how has the separation changed and what does it do to dose at the target?"Less tissue above the target raises dose there; a re-sim or recalc decision follows
In clinic

Weight loss in head and neck patients: a 1 cm loss of overlying tissue raises dose at a 6 MV target by roughly 3 percent (less attenuation), before any change in distance. That is part of why weekly imaging and mask fit matter.

13 · Rapid review

Board pearls and traps

PDD depends on SSD; TMR does not

TAR, TPR, TMR and SAR are all independent of SSD.

PDD rises with energy, field size and SSD

And falls with depth. Know the direction for every variable.

TMR at dmax = 1

By definition. TMR ≥ PDD/100 beyond dmax.

Sc uses jaws, Sp uses the blocked field

Sc,p = Sc × Sp. Sc is measured in air.

Equivalent square = 4A/P

Circle: 0.89 × diameter.

Upper jaw drives Sc

Collimator exchange effect.

Mayneord F > 1 when SSD increases

Corrects inverse square only; overstates the gain for large fields.

1 cm at 100 cm ≈ 2%

Inverse square near isocenter.

dmax: Co-60 0.5, 6 MV 1.5, 10 MV about 2.5, 18 MV about 3 to 3.5 cm

dmax moves slightly shallower with larger fields.

PDD(10) for 10 x 10: 6 MV ≈ 67%, 18 MV ≈ 79%

Co-60 ≈ 56% at 80 cm SSD.

TG-51 reference depth is 10 cm

Beam quality %dd(10)x; lead foil for 10 MV and above.

Isocentric ISF

(101.5/100)² = 1.030 at 6 MV for an SSD-calibrated machine; 1.00 if SAD-calibrated.

FFF beams

Higher dose rate, less head scatter, softer on axis, lower out-of-field dose.

Wedge angle for hinge θ

90° − θ/2.

Common traps
  • Using the field size at the surface for an isocentric Sp. Use the field at isocenter.
  • Forgetting the ISF of 1.030 when the machine is SSD-calibrated, or adding it when it is SAD-calibrated.
  • Dividing the total dose by MU per field without splitting it between the fields first.
  • Applying the F factor in the wrong direction. Extended SSD gives a higher PDD.
  • Putting a factor in the numerator. Every correction factor belongs in the denominator.
  • Collimator setting at extended SSD: the console setting is defined at 100 cm, not at the skin.
14 · Test yourself

Question bank

Exam-style single best answer. Pick one to see the explanation.

0 / 0 answered
15 · Keep this

Formula card and sources

SSD
MU = D ÷ [K · PDD/100 · Sc(rc) · Sp(r) · WF · TF · ISF]
ISF = (SCD / (SSD + dmax))²
SAD
MU = D ÷ [K · TMR · Sc(rc) · Sp(rd) · WF · TF · ISF]
ISF = (SCD / SAD)²
Inverse square
D2 = D1 · (r1 / r2)²
Mayneord F
F = [(f2+dm)/(f1+dm)]² · [(f1+d)/(f2+d)]²
Equivalent square
s = 4A/P = 2ab/(a+b)
circle: s ≈ 0.89 · D
Scatter
Sc,p = Sc · Sp
SAR(d,r) = TAR(d,r) − TAR(d,0)

Sources

  1. American Board of Radiology. Radiation Oncology Qualifying Exam: Study Guide for Medical Physics for Radiation Oncology (2023 domain ranges).
  2. Khan FM, Gibbons JP. The Physics of Radiation Therapy, 5th ed. Chapters on dose distribution, scatter analysis and dose calculation.
  3. Ford E. Primer on Radiation Oncology Physics. CRC Press, 2020.
  4. Almond PR et al. AAPM TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847.
  5. Stern RL et al. Verification of monitor unit calculations for non-IMRT clinical radiotherapy: AAPM TG-114. Med Phys 2011;38:504.
  6. Burmeister J et al. ASTRO core physics curriculum for radiation oncology residents. IJROBP 2015.

The interactive explorer and calculator use a simplified beam model fitted to typical published 6 and 18 MV behaviour. They teach direction and magnitude. Clinical numbers come from your machine's commissioning data.