The one idea
A monitor unit calculation answers one question: how much dose does one MU deliver to my point, in my patient, with my setup? Divide the prescribed dose by that number and you have the MU.
Think of calibration as a single known exchange rate: 1 MU buys 1 cGy, in water, at a fixed distance, at the depth of maximum dose, in a 10 x 10 cm field. Your patient differs from that reference in four ways, and each difference gets one correction factor:
- Distance from the source to the point. Corrected by the inverse square factor.
- Depth of tissue the beam crosses before the point. Corrected by PDD or TMR.
- Field size, which changes how much scattered radiation reaches the point. Corrected by Sc and Sp.
- Things in the beam such as wedges and block trays. Corrected by their transmission factors.
Every one of these factors sits in the denominator. If a factor is below 1 (deeper point, smaller field, a wedge), dose per MU falls, so you need more MU. If you remember nothing else, remember that direction. It lets you catch most calculation errors without finishing the arithmetic.
Before you calculate, predict. "Deep point, small field, wedge in: MU will be well above the dose in cGy." A prediction you make first protects you from a decimal slip.
Monitor units and calibration
An MU is not a dose. It is a count from the ionization chambers in the treatment head, called monitor chambers, which sit in the beam path below the flattening filter (or where it would be, in flattening-filter-free beams). The beam passes through them, they collect charge, and the machine stops when the count reaches the set MU. Physics then tunes the chamber electronics so that, under reference conditions, 1 MU produces 1 cGy.
There are two monitor chambers, a primary and a secondary. If the primary fails, the secondary terminates the beam a few MU later. A backup timer stands behind both. Modern chambers are sealed, so their reading does not drift with room temperature and pressure.
How the machine gets calibrated
Absolute calibration follows the AAPM TG-51 protocol. Physics places a calibrated ion chamber in a water tank at 10 cm depth, 10 x 10 cm field, at either 100 cm SSD or 100 cm SAD, measures charge, and converts it to dose to water using the chamber's calibration coefficient and a beam quality factor, kQ. Beam quality for photons is specified by %dd(10)x, the photon-only percent depth dose at 10 cm. The clinic then uses its own depth dose data to carry that measured dose back up to dmax, where the "1 cGy per MU" convention is usually defined.
TG-51 reference depth is 10 cm, not dmax, because at 10 cm the dose is free of the electron contamination that distorts the buildup region. Beam quality is %dd(10)x, measured at 100 cm SSD. For beams of 10 MV and above, a thin lead foil is placed in the beam during that measurement to account for contaminant electrons.
The convention that changes your arithmetic
Clinics define their reference point in one of two ways. You need to know which, because it decides the inverse square factor in an isocentric calculation.
| Convention | Reference point | Source to calibration point | ISF for an isocentric calc (6 MV) |
|---|---|---|---|
| SSD-calibrated | dmax, with the water surface at 100 cm | 100 + dmax = 101.5 cm | (101.5 / 100)² = 1.030 |
| SAD-calibrated | dmax, with that point at isocenter | 100 cm | (100 / 100)² = 1.000 |
Ask your physicist once: "Are our machines calibrated SSD or SAD, and at what depth?" Write it on your formula card. The answer changes every isocentric hand calc by about 3 percent at 6 MV and more at higher energy, because dmax is deeper.
Inverse square
The photon beam spreads from a nearly point-like source. The same number of photons passes through every cross-section of the beam, but the cross-section's area grows with the square of the distance. Photon fluence, and the dose it produces in air, therefore falls with the square of distance from the source.
Example: a point at 100 cm receives 100 cGy. At 110 cm it receives 100 × (100/110)² = 82.6 cGy. Ten percent farther away costs about 17 percent of the dose.
Inverse square describes the primary beam in air. Inside a patient, attenuation and scatter also act, which is why we need depth dose tables. Inverse square still appears inside every MU equation, but only to account for the change in distance between the calibration point and your point.
Near 100 cm, a 1 cm change in distance changes dose by about 2 percent: (100/101)² = 0.980. A patient set up 1.5 cm closer to the source than planned on an SSD treatment gets about 3 percent more dose at depth. This is the number behind every SSD check on day one.
Buildup and dmax
Photons do not deposit dose directly. They hand their energy to electrons (mostly by Compton scattering at therapy energies), and those electrons deposit dose along their tracks as they slow down. Megavoltage photons throw electrons mainly forward, and those electrons travel centimetres before stopping.
At the skin, few electrons have been set in motion yet, so dose is low. With each millimetre, electrons generated above add to those generated here. Dose rises until the number of electrons entering a small volume equals the number leaving it. That balance is called charged particle equilibrium, and the depth where dose peaks is dmax. Past dmax, the photon beam is being attenuated, fewer electrons are produced, and dose falls.
Because higher-energy electrons travel farther, dmax moves deeper as beam energy rises. That is the physical basis of skin sparing, and the reason high-energy beams spare skin better.
| Beam | dmax (approx.) | PDD at 10 cm, 10 x 10 | Clinical use |
|---|---|---|---|
| Co-60 | 0.5 cm | ≈ 56% (at 80 cm SSD) | Historical; Gamma Knife source |
| 4 MV | 1.0 cm | ≈ 62% | Older units, thin sites |
| 6 MV | 1.5 cm | ≈ 67% | Workhorse: HN, breast, extremity, most VMAT |
| 10 MV | 2.3 to 2.5 cm | ≈ 73% | Pelvis, thicker patients |
| 15 MV | ≈ 2.9 cm | ≈ 77% | Large separations |
| 18 MV | 3.0 to 3.5 cm | ≈ 79% | Large separations; produces neutrons |
Values are typical for 100 cm SSD unless noted. Your machine's commissioned data take precedence; these are the numbers to know for the exam.
What raises surface dose
Anything that adds electrons above the skin: larger fields (more electrons scattered from the head and air), a block tray or anything else close to the skin, bolus, and oblique incidence. Surface dose for a 6 MV 10 x 10 field at normal incidence is roughly 15 to 25 percent; it is lower at higher energy.
dmax shifts slightly shallower as field size increases, because larger fields carry more contaminating electrons from the treatment head. Bolus moves dmax up toward the skin by its thickness. Placing a tray close to the skin destroys skin sparing, which is why trays sit at least about 15 to 20 cm above the skin.
When you want full dose on a mastectomy scar or a superficial node, you are asking to defeat buildup. Bolus is the tool. When you want to protect skin in a deep pelvic target, buildup is your friend and higher energy helps.
Percent depth dose
Percent depth dose (PDD) is the dose at depth d on the central axis divided by the dose at dmax on the central axis, times 100, for one beam at a fixed source-to-surface distance (SSD). The point and the reference point are at different distances from the source. That detail is why PDD depends on SSD and why it is the natural quantity for SSD setups.
| Increase this | PDD beyond dmax | Why |
|---|---|---|
| Depth | Falls | Attenuation and inverse square both act |
| Energy | Rises | Higher-energy photons are less attenuated in water |
| Field size | Rises | More scatter reaches the deep point, relatively more than at dmax |
| SSD | Rises | The fractional distance between dmax and the deep point shrinks, so inverse square costs less |
The SSD effect is the one residents miss. At 100 cm SSD the dmax point is at 101.5 cm and a 10 cm point at 110 cm, an inverse square ratio of (101.5/110)² = 0.851. At 120 cm SSD the same depths sit at 121.5 and 130 cm, a ratio of 0.874. The deep point loses less to distance, so its PDD is higher. The absolute dose rate is lower at extended SSD, but the shape of the depth dose curve is more penetrating.
Depth dose explorer
Teaching model, not commissioning dataDashed line: the same energy at 10 x 10, 100 cm SSD, for reference. Move one slider at a time and say out loud why the curve moved.
Correcting PDD for a different SSD: the Mayneord F factor
PDD tables are measured at one SSD, usually 100 cm. For an extended SSD treatment (large fields, TBI, a patient who does not fit), you scale the table value by the Mayneord F factor, which corrects only the inverse square part of the change.
The F factor ignores the change in scatter. It works well for small fields and slightly overstates the PDD increase for large fields and deep points, where scatter makes up more of the dose. F is greater than 1 when SSD increases.
TAR, TPR and TMR
Modern treatments are isocentric. The machine rotates around a fixed point, the isocenter, 100 cm from the source, and we place the tumor there. From one beam to the next the point stays at 100 cm; what changes is how much tissue lies above it. PDD is clumsy here because it ties depth and distance together. We need a ratio where the point stays at the same distance from the source and only the depth of overlying tissue changes. Three such ratios exist.
In SSD geometry the point moves away from the source as depth increases, so PDD carries an inverse square fall-off. In SAD geometry the point never moves, so TMR carries attenuation and scatter only.
| Ratio | Definition (same point, same distance from source) | Where you meet it |
|---|---|---|
| TAR | Dose at depth d in phantom ÷ dose at the same point in free space (in air, with just enough buildup) | Co-60 era; awkward at MV because "in air" measurement needs a large buildup cap |
| BSF / PSF | TAR at dmax; how much phantom scatter adds at the peak. Called peak scatter factor at MV | Kilovoltage and orthovoltage, where it can exceed 1.3 |
| TPR | Dose at depth d ÷ dose at a reference depth, same point | General form; reference depth often 10 cm |
| TMR | TPR with the reference depth at dmax | Standard for isocentric hand calcs and many planning systems |
| SAR | TAR(d, r) − TAR(d, 0): the scatter part only | Clarkson integration for irregular fields |
TMR depends on energy, depth and field size at the point. It does not depend on SSD, because both doses in the ratio are measured at the same distance from the source, so inverse square cancels. That independence is exactly why it works for isocentric setups.
TMR is always at least as large as PDD at the same depth and field size beyond dmax, because PDD contains an extra inverse square fall-off. At 6 MV, 10 x 10, 10 cm deep: TMR ≈ 0.79 while PDD ≈ 67%.
Quantities independent of SSD: TAR, TPR, TMR, SAR. Quantity that depends on SSD: PDD. TMR at dmax equals 1 by definition. In Clarkson's method, the primary (zero-area TAR) and the scatter (SAR summed over radial sectors) are calculated separately and added.
Field size factors
Open the jaws and two things increase. More of the treatment head (mainly the flattening filter) is visible from the point, so more photons scatter out of the head toward it. And more patient tissue is irradiated, so more photons scatter within the patient toward it. The two are separated because blocking and MLC shaping change them differently.
| Factor | What it captures | Which field size to use | How it is measured |
|---|---|---|---|
| Sc collimator scatter, head scatter | Scatter from the head; output in air | Collimator (jaw) setting, defined at isocenter. Blocks and MLC do not change it (to first order) | In air, chamber with buildup cap or miniphantom, normalized to 10 x 10 |
| Sp phantom scatter | Scatter from the irradiated patient or water | Actual irradiated (blocked) field at the point of calculation | Derived: Sp = Sc,p ÷ Sc |
| Sc,p total scatter factor, output factor | Both together | Open field with no blocking | In water at reference depth, normalized to 10 x 10 |
A 20 x 20 jaw opening with MLC shaping down to an equivalent 12 x 12 field: Sc uses 20 x 20, Sp uses 12 x 12. Using 12 for both underestimates the head scatter and overstates the MU. Using 20 for both overstates the patient scatter and under-doses. Exams test this directly.
Rectangles and circles: equivalent square
Tables are indexed by square fields. A rectangular field gives the same central axis scatter as a square field of the same area-to-perimeter ratio (Sterling's rule).
Typical sizes of the effect
For a flattened 6 MV beam, Sc runs roughly from 0.96 at 4 x 4 to about 1.05 at 40 x 40; Sp spans a similar range. A field-size error of a few centimetres therefore moves dose by 1 to 3 percent. Small fields (below about 4 x 4) are a different regime: the source is partly hidden, lateral equilibrium is lost, and output drops steeply. That belongs to the SRS module.
Collimator exchange effect: a 5 x 20 field and a 20 x 5 field give slightly different Sc. The upper jaw, closer to the source, has more influence on head scatter, so Sc is larger when the upper jaw defines the long side. Sp is identical for both.
Flattening-filter-free beams have much less head scatter, so their Sc changes less with field size. They also deliver higher dose rates (on a TrueBeam, up to 1400 MU/min at 6 MV FFF and 2400 MU/min at 10 MV FFF, versus 600 MU/min flattened).
Beam modifiers
Anything placed in the beam is handled with a transmission factor: dose on the central axis with the device divided by dose without it, same MU. It sits in the denominator with everything else.
| Device | Typical factor | What else it changes |
|---|---|---|
| Block tray | ≈ 0.95 to 0.98 | Adds surface dose if close to skin |
| Physical wedge | ≈ 0.5 to 0.8, angle dependent | Hardens the beam, so PDD rises slightly; factor varies with depth and field size |
| Enhanced or dynamic wedge | varies with field length | A jaw sweeps during delivery. No hardening; factor depends strongly on field size along the sweep direction |
| MLC | built into Sp | Leakage and transmission of 1 to 2 percent through closed leaves; matters for heavily modulated plans |
The wedge angle is the angle by which an isodose line (classically the 50 percent line, or the line at 10 cm depth by ICRU) is tilted from perpendicular to the central axis. Hinge angle θ between two beams calls for a wedge angle of 90° − θ/2. Two beams at 90° need 45° wedges.
The two equations
Every hand calc on the exam is one of these two equations. Read them as the "one idea" from section 01 with every correction written out.
| K | Calibrated output, usually 1 cGy/MU at the reference point |
| SCD | Source to calibration point distance |
| rc | Collimator (jaw) setting at isocenter, as an equivalent square |
| rs | Blocked field size at the surface (SSD calc) |
| rd | Blocked field size at the point of calculation (SAD calc; at isocenter) |
| rdmax | Blocked field size at dmax; for SSD calcs at 100 cm, effectively the surface field |
| WF, TF | Wedge factor and tray factor; 1 if absent |
To find dose from MU, flip the equation: D = MU × K × (all the factors). Exams love "a field delivered X MU, what dose did point Y receive?" questions. The factors are identical; only the algebra moves.
MU calculator
This runs the equations above on a teaching model of a 6 MV and an 18 MV beam. It shows every factor, so you can change one input and watch which factor moves. Use it to check your own hand calcs from the worked examples; never use it for a patient.
Hand calc bench
Teaching model, not for clinical useJaw settings are collimator settings at 100 cm, as on the console.
Worked examples
Work each one on paper before opening the solution. Factors are given in the problem, as they are on the exam; the method is what you are practising.
Single field, fixed SSD
6 MV, 10 x 10 cm at the surface, 100 cm SSD. Prescribe 200 cGy at 5 cm depth. PDD(5) = 86.5%. Machine calibrated 1 cGy/MU at dmax, 100 cm SSD. No modifiers.
Show solution
Reference field, reference SSD: Sc = Sp = 1 and ISF = 1. Only depth changes.
≈ 231 MU
Sanity check: dose at dmax is 231 cGy, which is 200 ÷ 0.865. Every SSD field delivers more dose at dmax than at the prescription depth.
Single field, isocentric
6 MV, 10 x 10 cm at isocenter, isocenter 10 cm deep. Prescribe 200 cGy to isocenter. TMR(10, 10x10) = 0.790. Machine calibrated 1 cGy/MU at dmax with 100 cm SSD.
Show solution
The calibration point is at 101.5 cm, the isocenter at 100 cm. The isocenter is closer to the source, so it gets more dose per MU: ISF = (101.5/100)² = 1.030.
≈ 246 MU
Compare with an SSD setup to 10 cm depth: 200 ÷ 0.67 = 298.5 MU. The isocentric point is 10 cm closer to the source (100 versus 110 cm), so it needs about 18 percent fewer MU.
Parallel opposed, isocentric, blocked
AP/PA, 6 MV, isocenter at midplane of a 22 cm separation. 180 cGy per fraction to isocenter, equally weighted. Jaws 15 x 12, MLC blocks the field to an effective 12 x 10. Given: Sc(13.3) = 1.012, Sp(10.9) = 1.003, TMR(11, 10.9) = 0.770. SSD-calibrated machine. Find MU per field.
Show solution
Each field delivers half: 90 cGy at 11 cm depth. Collimator equivalent square: 2 × 15 × 12 / 27 = 13.3 cm, for Sc. Blocked equivalent square: 2 × 12 × 10 / 22 = 10.9 cm, for Sp and TMR.
≈ 112 MU per field
Both fields get the same MU only because isocenter is at midplane. If the isocenter were off midline, the AP and PA depths differ and so do their MU.
Extended SSD
6 MV single field at 120 cm SSD. Field is 10 x 10 at the surface. Prescribe 200 cGy at 10 cm depth. PDD(10, 10x10) at 100 cm SSD = 67.0%. Sc(8.3) = 0.992, Sp ≈ 1.00. SSD-calibrated machine.
Show solution
Step 1, collimator setting: a 10 cm field at 120 cm was set as 10 × 100/120 = 8.3 cm at isocenter, so Sc(8.3) = 0.992.
Step 2, PDD at the new SSD: F = (121.5/101.5)² × (110/130)² = 1.026, so PDD = 67.0 × 1.026 = 68.7%.
Step 3, distance: the dmax point is now at 121.5 cm versus 101.5 cm at calibration. ISF = (101.5/121.5)² = 0.698.
≈ 421 MU
Extended SSD costs a lot of MU because inverse square dominates, even though the PDD improves slightly. That is the trade at the heart of TBI.
Add a wedge
Repeat EX 2 with a 45° physical wedge, wedge factor 0.72.
Show solution
≈ 341 MU
A wedge factor of 0.72 raises MU by 1/0.72 = 1.39 times. If a wedged field shows the same MU as the open field, the wedge was left out of the calculation.
Reverse: what did the skin side get?
A single 6 MV SSD field (100 cm SSD, 10 x 10) is prescribed 200 cGy at 10 cm depth. PDD(10) = 67%. What is the dose at dmax?
Show solution
≈ 299 cGy, about 149% of the prescription
This is why single deep fields are rarely used: the entrance region is far hotter than the target. Opposed fields, multiple fields and higher energy all exist to fix that.
In the planning room
You will not hand-calculate a VMAT plan, but the same physics decides whether a plan's MU look reasonable, whether a setup discrepancy matters, and what to ask when a number looks wrong.
Ballpark numbers to carry in your head
| Situation | Rough number | What it tells you |
|---|---|---|
| 6 MV, 10 x 10, 200 cGy to isocenter at 10 cm | ≈ 245 MU | Dose per MU near 0.8 cGy at typical depths |
| Same, 18 MV | ≈ 210 MU | Higher energy: fewer MU for the same deep dose |
| Plan MU far above dose in cGy for 3D fields | MU/cGy > 1.5 | Deep point, small field, wedge, or an error. Find which. |
| VMAT or IMRT versus 3D for the same target | often 1.5 to 3 times the MU | Modulation. Very high MU means more leakage dose and more sensitivity to MLC errors |
| SSD off by 1 cm | ≈ 2% at depth | Distance errors matter at the percent level |
| Treatment time | MU ÷ dose rate | 300 MU at 600 MU/min is 30 s of beam-on |
The independent MU check
Every plan gets a second, independent calculation of MU, by hand or with separate software, before first treatment (AAPM TG-114 covers this). For simple geometry the two should agree within a few percent; heterogeneous anatomy, small or highly blocked fields, points far off-axis and oblique surfaces justify wider tolerances, and each clinic sets its own action level. IMRT and VMAT are verified differently, through patient-specific QA measurement or an independent dose recalculation, because a single point calculation cannot describe a modulated field.
When the planning system and the check disagree, the usual reasons are predictable: lung or bone heterogeneity (the hand calc assumes water), a point near a field edge or in a steep gradient, missing tissue or oblique incidence, a wrong field size in Sc or Sp, or a forgotten wedge or tray.
| Say this to physics or dosimetry | Why it lands |
|---|---|
| "What's the second-check agreement on this plan, and where's the calc point?" | A good agreement at a bad point (in a gradient or in lung) means little |
| "Is the calc point in lung? How much of the difference is heterogeneity correction?" | Shows you know a water-based check will disagree in lung, and by roughly how much |
| "The MU look high for this depth. Is there a wedge, or is the field small?" | Uses the direction-of-factors logic from section 01 |
| "How many MU per Gy on this VMAT? Can we lower modulation without losing coverage?" | Lower MU means less leakage and a plan more robust to small MLC errors |
| "The patient lost weight; how has the separation changed and what does it do to dose at the target?" | Less tissue above the target raises dose there; a re-sim or recalc decision follows |
Weight loss in head and neck patients: a 1 cm loss of overlying tissue raises dose at a 6 MV target by roughly 3 percent (less attenuation), before any change in distance. That is part of why weekly imaging and mask fit matter.
Board pearls and traps
TAR, TPR, TMR and SAR are all independent of SSD.
And falls with depth. Know the direction for every variable.
By definition. TMR ≥ PDD/100 beyond dmax.
Sc,p = Sc × Sp. Sc is measured in air.
Circle: 0.89 × diameter.
Collimator exchange effect.
Corrects inverse square only; overstates the gain for large fields.
Inverse square near isocenter.
dmax moves slightly shallower with larger fields.
Co-60 ≈ 56% at 80 cm SSD.
Beam quality %dd(10)x; lead foil for 10 MV and above.
(101.5/100)² = 1.030 at 6 MV for an SSD-calibrated machine; 1.00 if SAD-calibrated.
Higher dose rate, less head scatter, softer on axis, lower out-of-field dose.
90° − θ/2.
- Using the field size at the surface for an isocentric Sp. Use the field at isocenter.
- Forgetting the ISF of 1.030 when the machine is SSD-calibrated, or adding it when it is SAD-calibrated.
- Dividing the total dose by MU per field without splitting it between the fields first.
- Applying the F factor in the wrong direction. Extended SSD gives a higher PDD.
- Putting a factor in the numerator. Every correction factor belongs in the denominator.
- Collimator setting at extended SSD: the console setting is defined at 100 cm, not at the skin.
Question bank
Exam-style single best answer. Pick one to see the explanation.
Formula card and sources
Sources
- American Board of Radiology. Radiation Oncology Qualifying Exam: Study Guide for Medical Physics for Radiation Oncology (2023 domain ranges).
- Khan FM, Gibbons JP. The Physics of Radiation Therapy, 5th ed. Chapters on dose distribution, scatter analysis and dose calculation.
- Ford E. Primer on Radiation Oncology Physics. CRC Press, 2020.
- Almond PR et al. AAPM TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847.
- Stern RL et al. Verification of monitor unit calculations for non-IMRT clinical radiotherapy: AAPM TG-114. Med Phys 2011;38:504.
- Burmeister J et al. ASTRO core physics curriculum for radiation oncology residents. IJROBP 2015.
The interactive explorer and calculator use a simplified beam model fitted to typical published 6 and 18 MV behaviour. They teach direction and magnitude. Clinical numbers come from your machine's commissioning data.