Module 06 · ABR domain II.B · Radiation measurements

Detectors and TG-51 Calibration

The "1 cGy per MU" printed on every plan is a measured number. Once a year a physicist puts a calibrated ion chamber in a water tank, collects a few nanocoulombs, and applies a short chain of corrections to turn that charge into dose to water. Every patient's dose inherits the accuracy of that chain. This module builds the chain factor by factor, then covers the other detectors you will meet: the diode on a patient's skin, the film behind an IMRT QA, the OSLD in a pacemaker check, and the survey meter at the door of an HDR suite.

In this module
  1. 01The one idea
  2. 02Gas detectors and the voltage curve
  3. 03Ion chambers and electrometers
  4. 04Correcting the reading
  5. 05TG-51 for photons
  6. 06TG-51 for electrons
  7. 07Traceability and output checks
  8. 08Solid-state and film detectors
  9. 09Survey meters and neutron detectors
  10. 10TG-51 worksheet
  11. 11Worked examples
  12. 12In the planning room
  13. 13Board pearls and traps
  14. 14Question bank
  15. 15Formula card and sources
01 · Start here

The one idea

A detector reports a signal, never a dose. Calibration is the short, written chain of factors that turns one specific chamber's charge, in one specific beam, under one day's air conditions, into dose to water.

DwQ = M · kQ · ND,w60Co    M = Pion · PTP · Pelec · Ppol · Pleak · Mraw
Mraw is the raw charge. The P factors fix the reading for incomplete collection, air density, the electrometer, cable and polarity effects, and leakage. ND,w is the chamber's dose-to-water calibration coefficient in a cobalt beam, from an accredited lab. kQ carries that coefficient from cobalt to your beam.

Module 05 showed how a small air cavity reports dose to the water around it: charge over air mass, times W/e, times a stopping power ratio. TG-51, the AAPM protocol for reference dosimetry of megavoltage photon and electron beams, keeps that physics but hides most of it. The lab measures the chamber in cobalt directly in dose to water, so W/e and the stopping power ratio never appear explicitly. What remains for the clinic is to correct the reading and to apply kQ, the ratio of the chamber's dose-to-water response in cobalt to its response in the clinical beam.

Three answers carry the module:

  1. Every P factor corrects the reading toward what an ideal chamber at reference air conditions would have collected. Most sit within a percent of 1.000; PTP can move by several percent.
  2. Beam quality picks kQ. Photons use %dd(10)x; electrons use R50.
  3. Different jobs need different detectors. Ion chambers for reference dose, diodes and diamonds for small fields and in vivo checks, film for planar distributions, survey meters for protection.
Pearl

Read every correction factor as a question: did the chamber collect too little (factor above 1) or too much (factor below 1) compared with an ideal chamber at 22 °C and 101.33 kPa?

02 · One gas, six behaviours

Gas detectors and the voltage curve

Every gas-filled detector is the same object: two electrodes with gas between them and a voltage across it. Radiation frees ion pairs in the gas, and the field drags them to the electrodes. What changes from one instrument to the next is only the voltage, and that changes what happens to the ions on the way.

At very low voltage, ions drift slowly and many recombine before they arrive, so the signal is incomplete. Raise the voltage and collection becomes complete: every ion pair formed is collected and no more. That plateau is the ionization chamber region, and it is where dosimetry lives, because the charge is proportional to the energy deposited. Raise it further and the freed electrons gain enough energy between collisions to ionize the gas themselves, multiplying the charge by a fixed gain: the proportional region. Higher still, each event triggers an avalanche along the whole anode and every pulse is the same size, regardless of what started it: the Geiger-Müller region. Beyond that, the gas discharges continuously and the detector is useless.

I II III IV V VI large event small event all pulses equal Applied voltage → log collected charge per event →

Schematic, not to scale. Amber: an event that frees many ion pairs (an alpha particle). Blue: one that frees few (a beta particle). The two stay separate through the proportional region and merge in the Geiger-Müller region, where pulse size no longer carries any information about the event. Shaded: the ionization chamber plateau.

RegionWhat happens to the ionsInstrumentUse
I RecombinationSome recombine before collectionNoneAvoid; this is what Pion corrects at the edge of the plateau
II Ionization chamberAll collected, no multiplicationFarmer and parallel-plate chambers, well chambers, ion chamber survey metersDosimetry, calibration, accurate exposure rate
III ProportionalGas multiplication by a fixed gain; pulse size tracks energy depositedProportional counters, BF3 and He-3 neutron countersNeutron detection, telling alpha from beta
IV Limited proportionalGain starts to saturateNone in routine useAvoid
V Geiger-MüllerFull avalanche every event; identical pulsesGM countersFinding contamination and low-level radiation; counting, not dosimetry
VI Continuous dischargeSpontaneous dischargeNoneDamages the tube
Board point

Ion chambers are the dosimetry standard because in region II the collected charge equals the charge produced, with no gain to drift. A GM counter is the most sensitive gas detector but gives the same pulse for every event, so it cannot measure energy or dose without energy compensation and calibration.

03 · The reference instrument

Ion chambers and electrometers

A clinical ion chamber is a small, open-to-air cavity with a thin conducting wall and a central collecting electrode, connected by a triaxial cable to an electrometer, a very sensitive charge meter. A Farmer chamber holds about 0.6 cm³ of air and collects roughly 20 nC per gray. The bias voltage, typically 300 V or less, is chosen to sit on the plateau of region II.

Chamber shape follows the job. A cylindrical (thimble) chamber is robust and nearly isotropic, so it is the reference chamber for photons. A parallel-plate chamber has a thin entrance window and a narrow gap, so it can measure close to a surface and in steep gradients: buildup region, low-energy electron beams. A well chamber surrounds a brachytherapy source and measures its air kerma strength (Module 15). Small-volume chambers (0.01 to 0.1 cm³) reduce volume averaging in small fields at the cost of signal.

Where the chamber "measures"

A cylindrical chamber replaces a cylinder of water with air. Electrons reaching the cavity come mostly through its upstream wall, so the chamber reads as if it sat slightly upstream of its centre. For depth-dose curves, physics shifts the curve by 0.6 rcav upstream for photons and 0.5 rcav for electrons, where rcav is the cavity radius. For a parallel-plate chamber, the point of measurement is the inner face of the entrance window. In TG-51 photon reference dosimetry, the chamber centre goes at 10 cm and no shift is applied, because kQ already contains the gradient effect.

EffectWhat causes itHow it shows up
LeakageCurrent through cable and insulators with no beam onDrift of reading; must be under 0.1% of signal for a reference-class chamber
Stem and cable signalIrradiated stem or cable produces its own chargeField-size dependent error; part of the polarity effect
SettlingInsulator charge redistributes after voltage is appliedReading drifts for the first minutes; pre-irradiate before measuring
Volume averagingCavity larger than the uniform part of the fieldReads low on the peak of a small field or steep gradient
Fluence perturbationAir cavity changes the electron fluence it measuresBuilt into kQ for reference conditions
In clinic

A 0.6 cm³ Farmer chamber is about 2.3 cm long. Put it in a 2 × 2 cm SRS field and it averages over the penumbra and reads several percent low. Small-field output factors are measured with small chambers, diodes, diamond or scintillator detectors and corrected per TRS-483 (Module 17).

04 · From raw charge to M

Correcting the reading

Temperature and pressure, PTP

The chamber is open to the room, so the mass of air in it changes with the weather. Warmer air is less dense, so fewer ion pairs form for the same dose and the raw reading is low. Lower pressure does the same. PTP scales the reading to the air density at TG-51 reference conditions, 22 °C and 101.33 kPa (760 mmHg). Its shape is the ideal gas law: density goes as P/T, so the correction goes as T/P.

PTP = 273.2 + T295.2 × 101.33P
T in °C, P in kPa. A hot day or a high-altitude clinic gives PTP > 1. Each 1 °C is about 0.34%; each 1 kPa about 1%. Denver sits near 83 kPa, which makes PTP about 1.22.

Ion recombination, Pion

Even on the plateau a small fraction of ions recombine. Initial recombination happens within one particle's track. General recombination happens between ions from different tracks, and it grows with the density of ions, which in a pulsed linac means dose per pulse. Raising the voltage collects faster and loses less, so the loss can be measured by reading at two voltages. For a pulsed beam, with VH the normal voltage and VL at most half of it:

Pion(VH) = 1 − VH/VLMrawH/MrawL − VH/VL   (pulsed)
For continuous beams (Co-60) the voltage ratios are squared. Pion is always ≥ 1. TG-51 says a chamber with Pion above 1.05 should not be used; typical values are 1.002 to 1.01, higher in FFF beams because they deliver more dose per pulse.

Polarity, Ppol

Reversing the bias should give the same magnitude of charge. It does not quite, because the beam itself deposits or knocks out charge in the collector, stem and cable, and that stray charge adds with one polarity and subtracts with the other. Averaging the two readings removes it.

Ppol = |Mraw+ − Mraw−|2 |Mraw|
Mraw in the denominator is the reading at the polarity used for the calibration. The TG-51 photon addendum expects 0.996 to 1.004 for a reference-class chamber. The effect is largest in electron beams, with parallel-plate and small chambers, and with a lot of cable in the field.

Electrometer and leakage

Pelec is the electrometer's own calibration factor, set to 1.000 when the lab calibrates chamber and electrometer as one unit. Pleak corrects for charge collected with no beam; the photon addendum asks that leakage stay under 0.1% of the reading.

ConditionFactorDirectionWhy
Room warmer than 22 °CPTP> 1Less air mass in the cavity, raw reading low
Pressure below 101.33 kPa (altitude, storm)PTP> 1Less air mass
Higher dose per pulse (FFF, short SSD)PionRisesMore general recombination
Higher bias voltagePionFalls toward 1Faster collection, less recombination
Electron beam, parallel-plate chamberPpolFarther from 1More charge deposited in collector and stem
Trap

Writing PTP upside down. A hot day means less air and a low raw reading, so the correction must be above 1. If your factor came out below 1 for a day warmer than 22 °C at normal pressure, it is inverted.

05 · Reference dosimetry

TG-51 for photons

TG-51 (Almond et al., 1999), updated for photons by the 2014 addendum (McEwen et al.), defines one measurement: dose to water at 10 cm depth on the central axis, 10 × 10 cm field, at 100 cm SSD or 100 cm SAD. Ten centimetres is chosen because contaminant electrons from the head have died out there, while the chamber is still deep in a region with a gentle gradient.

Why kQ exists

The chamber's calibration coefficient is measured in cobalt-60. In a 6 or 18 MV beam the water-to-air stopping power ratio is slightly smaller and the chamber's wall and cavity perturb the fluence slightly differently. kQ is the ratio of the chamber's calibration coefficient in your beam to its coefficient in cobalt. It is below 1 and falls as energy rises, by a few percent at 18 MV. It is chamber specific, which is why the protocol tabulates it per chamber model.

Beam quality: %dd(10)x

To look up kQ you need a number that describes the beam's penetration. TG-51 uses %dd(10)x: the percent depth dose at 10 cm for a 10 × 10 cm field at 100 cm SSD, for photons only. The subscript x matters. At 10 MV and above, electrons from the head add dose near dmax, which raises the denominator and lowers the measured %dd(10). To remove their effect, the depth dose is measured with a 1 mm lead foil in the beam, which replaces the head's contaminant electrons with a smaller, known contribution, and then converted.

%dd(10)x = %dd(10)   (below about 10 MV, %dd(10) < 75%)
%dd(10)x = [0.8905 + 0.00150 · %dd(10)Pb] · %dd(10)Pb   (foil 50 cm from surface, %dd(10)Pb ≥ 73%)
%dd(10)x = [0.8116 + 0.00264 · %dd(10)Pb] · %dd(10)Pb   (foil 30 cm from surface, %dd(10)Pb ≥ 71%)
%dd(10)x = 1.267 · %dd(10) − 20.0   (no foil, flattened beams, 75% < %dd(10) ≤ 89%)
The foil is removed for the dose measurement itself. The addendum allows the no-foil formula only for flattened beams, with a small added uncertainty, and recommends the foil for all FFF beams, including those below 10 MV.

With %dd(10)x in hand, kQ for a reference-class chamber comes from the addendum's fit, valid from 63% to 86%:

kQ = A + B·10−3·%dd(10)x + C·10−5·(%dd(10)x)2
Dw(10 cm) = M · kQ · ND,w   →   D(dmax) = Dw(10 cm) / [%dd(10)/100]  (SSD setup)
A, B, C are chamber-specific. Below 63% interpolate linearly to kQ = 1.000 at cobalt (%dd(10)x = 58). For an SAD setup, the clinic carries the 10 cm dose to its normalization point with TMR instead of PDD. Note that the dose transfer uses the clinical %dd(10), not %dd(10)x.
ChamberABCkQ at 66.4% (6 MV)kQ at 80.4% (18 MV)
NE2571 (0.6 cm³ Farmer)0.98821.486−2.1400.99250.9693
Exradin A12 (Farmer type)1.01460.777−1.6660.99270.9694
PTW 30012 (0.6 cm³ Farmer)0.95372.440−2.7500.99450.9721

Coefficients from Table I of the TG-51 photon addendum (McEwen et al., 2014). The physicist uses the values for the exact chamber model in use.

Board point

TG-51 photon reference conditions: 10 cm depth, 10 × 10 cm, 100 cm SSD or SAD, chamber centre at 10 cm. Beam quality: %dd(10)x at 100 cm SSD. Lead foil for %dd(10) of about 75% and above (10 MV and up), and for all FFF beams. kQ < 1 and falls with energy. The international equivalent, IAEA TRS-398, uses TPR20,10 as its beam quality index instead.

06 · Reference dosimetry

TG-51 for electrons

Electron depth doses have no long exponential tail to sit in. The dose rises, peaks, and falls steeply to the practical range within a few centimetres, and the electron spectrum changes with depth, so the stopping power ratio changes too. TG-51 handles this by measuring at a depth tied to the beam's own range.

Beam quality: R50

R50 is the depth in water where dose falls to 50% of its maximum, measured at 100 cm SSD with a field of at least 10 × 10 cm (20 × 20 cm when R50 exceeds 8.5 cm). An ion chamber measures ionization, not dose, so physics first finds I50, the depth of 50% ionization, and converts:

R50 = 1.029 · I50 − 0.06 cm   (2 ≤ I50 ≤ 10 cm)
R50 = 1.059 · I50 − 0.37 cm   (I50 > 10 cm)
dref = 0.6 · R50 − 0.1 cm
dref sits close to dmax for low energies and somewhat deeper than dmax for high energies. At this depth the stopping power ratio is described well by R50 alone. Mean incident energy is roughly 2.33 × R50 MeV (Module 09).

Original protocol (1999)

The 1999 protocol split kQ into three pieces. kecal is a fixed, chamber-specific photon-to-electron conversion at one reference electron quality. k′R50 carries it to the actual R50. PgrQ is a gradient correction for cylindrical chambers, measured as the ratio of readings with the chamber centre at dref + 0.5 rcav and at dref. Parallel-plate chambers were required for low energies (R50 ≤ 4.3 cm, about 10 MeV and below) and need no gradient correction.

Dw(dref) = M · PgrQ · k′R50 · kecal · ND,w60Co
k′R50 (cylindrical) = 0.9905 + 0.0710 · e−R50/3.67   (2 ≤ R50 ≤ 9 cm)
Pgr is below 1 for cylindrical chambers because the reading rises toward the surface on the upslope of an electron depth dose. Dose at dmax = Dw(dref) divided by the clinical PDD at dref.

2024 electron addendum

The AAPM working group on TG-51 published an electron addendum in 2024 (Medical Physics, WGTG51 Report 385). It allows calibrated reference-class cylindrical chambers in all electron beams, 4 to 22 MeV, with no separate gradient correction: the gradient effect is folded into a single Monte Carlo-based kQ that depends on R50. It reports that dose per MU may come out up to about 2% higher than with the 1999 protocol. Learn the 1999 structure for the exam, because older questions use it, and know that your clinic may now calibrate electrons with a Farmer chamber at every energy.

ItemPhotonsElectrons
Beam quality%dd(10)xR50 (from I50)
Reference depth10 cm0.6 R50 − 0.1 cm
Field and distance10 × 10 cm, 100 cm SSD or SAD≥ 10 × 10 cm (≥ 20 × 20 if R50 > 8.5 cm), 100 cm SSD
ChamberCylindrical, reference class1999: parallel-plate required at low energy. 2024: calibrated cylindrical allowed at all energies
GradientInside kQ1999: Pgr. 2024: inside kQ
Point of measurement for depth curvesShift 0.6 rcav upstreamShift 0.5 rcav upstream
Trap

Using I50 as if it were R50, or setting dref = 0.6 R50 and forgetting the −0.1 cm. Both shift the measurement depth by about a millimetre, which on the steep part of a low-energy electron curve is not trivial.

07 · The chain of custody

Traceability and output checks

Your chamber's ND,w means something only because it ties back to a primary standard. In the United States, NIST realizes absorbed dose to water in a cobalt beam with a water calorimeter: a gray of dose raises the temperature of water by only about 0.24 mK, and that temperature rise, with the specific heat of water, is dose by definition. NIST calibrates the reference chambers of the Accredited Dosimetry Calibration Laboratories (ADCLs). The ADCL calibrates your chamber and electrometer in its cobalt beam. TG-51 then carries the coefficient to your linac.

LinkWhoWhat they establish
Primary standardNISTDose to water in Co-60 by calorimetry; air kerma by graphite cavity chambers; kV x-ray air kerma by free-air chambers
Secondary standardADCL (AAPM accredited)ND,w60Co for the clinic's chamber and electrometer
Reference dosimetryClinic physicist, TG-51Dose per MU under reference conditions, at least annually
Output constancyTherapists and physics, TG-142Daily and monthly checks against the annual value
Independent auditIROC HoustonMailed OSLD output audits and phantom tests for trial credentialing

TG-51 requires the chamber and electrometer to have been calibrated within the previous two years. With reference-class equipment and full care, the photon addendum puts the combined standard uncertainty in dose at the reference point at about 0.9%; with questionable equipment and typical shortcuts it is about 2%.

Between annual calibrations, TG-142 sets output tolerances: 3% on daily constancy checks, 2% on monthly checks against the baseline, and 1% for the annual reference calibration. A daily device that reads 3.5% high is a reason to hold treatment and call physics, not to treat and note it.

Board point

The primary standard for absorbed dose to water is calorimetry. Fricke ferrous sulfate is a chemical dosimeter with a known radiation yield and has served as a reference. Free-air ionization chambers are the primary standard for kilovoltage air kerma, not for megavoltage beams, because megavoltage secondary electrons travel too far to be stopped in a practical chamber.

In clinic

A trial that requires IROC credentialing will not open at your site until a mailed OSLD audit and, for some protocols, an anthropomorphic phantom pass. If the IROC output result differs from your stated value by more than its tolerance, physics will be asked to recheck the TG-51 chain before any patient is enrolled.

08 · Everything that is not an ion chamber

Solid-state and film detectors

Ion chambers are accurate because air is predictable, but air is thin and gives little signal per volume. Solid detectors trade some of that predictability for size and sensitivity. Each one fails in a characteristic way, and the failure follows from its material.

Diodes

A silicon diode is a solid-state ion chamber. Radiation frees electron-hole pairs in the depletion region and the built-in field collects them with no bias voltage. Silicon is nearly 2000 times denser than air and needs only about 3.6 eV per pair instead of 34 eV, so per unit volume a diode gives roughly 18,000 times the signal of an air chamber. That makes it tiny and fast, ideal for small fields, scanning and in vivo checks. Its weaknesses come from silicon itself: Z = 14 over-responds to low-energy scatter; sensitivity changes with temperature (a few tenths of a percent per °C), with dose per pulse (so with SSD), with beam angle, and falls with accumulated dose as the crystal is damaged. In vivo diodes carry a buildup cap, so they report dose near dmax, not at the skin.

MOSFETs

A MOSFET stores charge in its oxide layer, which shifts its threshold voltage in proportion to dose. It is very small, reads out immediately, and is used in vivo, but it has a finite lifetime and modest precision.

TLD and OSLD

Both are crystals that trap electrons in defect states and release them, with light, when stimulated. TLD (thermoluminescent dosimeter, usually LiF:Mg,Ti, effective Z about 8.2) is heated; the light versus temperature is the glow curve. LiF is close to tissue, becomes supralinear above roughly 1 Gy, and reading it erases most of the signal. OSLD (optically stimulated luminescence, Al2O3:C, effective Z about 11.3) is stimulated with light instead of heat, can be re-read because each readout takes only a small fraction of the signal, and over-responds several-fold to kilovoltage photons because of its higher Z. OSLDs are the badge dosimeters for staff and the mailed audit dosimeters for IROC.

Film

Radiographic film (silver halide) needs chemical processing and over-responds strongly at low energy. Radiochromic film (for example EBT3 or EBT4) darkens by polymerization with no processing, is close to tissue equivalent, has fine spatial resolution, and is read on a flatbed scanner after the colour has stabilized, usually about a day. It is the standard for planar dose checks and small-field profiles; its weaknesses are scanner non-uniformity and orientation dependence.

Diamond and scintillators

Synthetic diamond detectors (carbon, Z = 6) are near water equivalent, very small and radiation hard, and are favoured for small-field output and profiles. Plastic scintillators are the most water-equivalent detectors in use: the plastic matches water in electron density and stopping power, so they barely perturb the field and show almost no energy dependence. Their one artifact is Cherenkov light produced in the optical fibre, which must be subtracted.

DetectorStrengthMain weaknessTypical use
Farmer ion chamberAccurate, stable, traceableVolume averaging in small fieldsTG-51, output checks
Parallel-plate chamberThin window, small gapPolarity effectElectrons, buildup, surface dose
DiodeTiny, sensitive, no biasTemperature, dose rate, angle, energy, damageScanning, small fields, in vivo
MOSFETTiny, immediate readoutFinite life, precisionIn vivo
TLD (LiF)Near tissue, passive, any shapeSupralinearity, erased on readout, slowIn vivo, audits, personnel
OSLD (Al2O3:C)Re-readable, passiveOver-responds at kVPersonnel badges, IROC audits, in vivo
Radiochromic film2D, high resolution, near tissueScanner handling, delay to readPlanar QA, small-field profiles
DiamondSmall, near water, robustCost; some dose-rate dependenceSmall-field output and profiles
Plastic scintillatorMost water-equivalent, real timeCherenkov stem signalSmall fields, in vivo research
Board point

Over-response at kilovoltage energies follows effective Z: silicon diodes, OSLD and radiographic film over-respond; LiF TLD slightly; radiochromic film, diamond and plastic scintillators are closest to water. Only an ionization chamber calibrated in dose to water is used for reference dosimetry.

09 · Protection instruments

Survey meters and neutron detectors

A survey meter answers a different question from a dosimeter: is there radiation here, and how much per hour? The choice depends on whether you need sensitivity or accuracy.

A Geiger-Müller counter works in region V. One ion pair triggers a full avalanche, so it detects almost anything, which makes it the tool for finding contamination, a lost seed, or a leak. Every pulse is the same, so it measures counts, not energy, and its exposure-rate reading is reliable only for the energy it was calibrated at. After each pulse the tube is dead for roughly 100 μs while the avalanche clears. In a very intense field the tube spends all its time dead and can read low or even zero; this is paralysis, and it is the reason a GM counter is never the only meter used near a high-activity source.

An ion chamber survey meter works in region II with a large, often pressurized air volume. It is far less sensitive, but its response is nearly flat with energy and it reads exposure rate accurately in high fields. It is the instrument for shielding surveys of a linac vault and for measuring dose rate at 1 m from a patient before release.

A scintillation detector (for example NaI(Tl)) converts gamma energy to light. It is the most sensitive photon survey instrument and can identify isotopes by energy.

Neutrons carry no charge and ionize nothing directly. A neutron detector therefore surrounds a proportional counter filled with BF3 or He-3 with polyethylene. The hydrogen in the moderator slows fast neutrons to thermal energies, where boron-10 or helium-3 captures them and releases charged particles that the counter sees. Shaped correctly, the moderator makes the response follow dose equivalent, which is why this is called a rem meter. Bubble detectors are passive alternatives. Neutron surveys are needed around vault doors and mazes for beams above about 10 MV (Module 21).

InstrumentRegionStrengthFailure modeUse it for
GM counterVHighest sensitivity of gas detectorsParalysis in high fields; energy dependentContamination, lost seeds, post-procedure room survey
Ion chamber survey meterIIAccurate exposure rate, flat energy responseLow sensitivity, slow responseVault shielding surveys, patient release
NaI scintillatorn/aVery sensitive, isotope identificationEnergy dependent, fragileLow-level photon surveys, spectroscopy
BF3 or He-3 rem meterIIINeutron dose equivalentHeavy; poor response to some energiesHigh-energy linac and proton vault surveys
Board point

Under 10 CFR 35.61, survey instruments used for licensed material are calibrated before first use, annually, and after any repair that could affect calibration. A GM counter can read near zero in a very high field because of dead time.

In clinic

After an HDR treatment, the therapist surveys the patient and the room with a meter to confirm the source has retracted. After an LDR prostate implant, the room and linens are surveyed with a GM or scintillation meter to find any stray seed. Both habits exist because a stuck source or a lost seed is invisible without an instrument.

10 · Run it

TG-51 worksheet

This is the photon calibration your physicist runs once a year, with every factor visible. The defaults are a plausible 6 MV calibration at 100 cm SSD. Change the room conditions, the chamber or the readings and watch which factors move and how much the final cGy per MU moves with them.

Raw charge to cGy per MU

Teaching model, not a calibration
Readings (nC)

For 6 MV, %dd(10)x equals the clinical %dd(10). For 10 MV and above, set %dd(10)x from the lead-foil measurement and keep the clinical %dd(10) for the dose transfer.

0cGy per MU at dmax
kQ0
PTP0
Pion0
Ppol0
Dose at 10 cm0
Off 1.000 cGy/MU0

Model: TG-51 photon formalism with the addendum kQ fit, pulsed-beam two-voltage Pion, and Pleak taken as 1. Default readings and ND,w are invented for teaching. The 2% and 1% flags follow TG-142 monthly and annual output tolerances.

11 · Practice the method

Worked examples

Work each one on paper first. The constants are given in the problem, as they are on the exam.

EX 1

Temperature and pressure correction

An output check is done in a room at 24.0 °C and 99.0 kPa with a vented chamber. Reference conditions are 22 °C and 101.33 kPa. Find PTP.

Show solution
PTP = (273.2 + 24.0)/295.2 × 101.33/99.0 = 1.0068 × 1.0235 = 1.0305

PTP ≈ 1.030

Sanity check: the room is warmer and the pressure lower than reference, so there is less air in the cavity and the raw reading is low; the factor must be above 1. The inverted factor, 0.970, would put the reported output 6% below the corrected one. Using pressure alone gives 1.024.

EX 2

Recombination and polarity

In a pulsed 6 MV beam a chamber reads 20.12 nC at +300 V, 20.03 nC at +150 V, and −20.16 nC at −300 V. Calibration is done at +300 V. Find Pion and Ppol.

Show solution
VH/VL = 2;   MH/ML = 20.12/20.03 = 1.00449
Pion = (1 − 2)/(1.00449 − 2) = 1.0045
Ppol = |20.12 − (−20.16)| / (2 × 20.12) = 1.0010

Pion ≈ 1.0045; Ppol ≈ 1.0010

Sanity check: Pion must exceed 1, since a higher voltage always collects at least as much charge, and here it is well under the 1.05 limit. Putting the low-voltage reading on top (20.03/20.12) gives 0.9955, which is impossible for a recombination correction. Ppol is inside the 0.996 to 1.004 window.

EX 3

Full photon calibration

A 6 MV beam is calibrated at 100 cm SSD, 10 × 10 cm, chamber centre at 10 cm. %dd(10) = %dd(10)x = 66.4%. NE2571 chamber, ND,w = 4.50 × 107 Gy/C, kQ from the addendum fit (A = 0.9882, B = 1.486, C = −2.140). For 100 MU: +300 V reads 14.70 nC, +150 V reads 14.67 nC, −300 V reads −14.72 nC. T = 21.0 °C, P = 101.8 kPa, Pelec = 1.000. Find dose per MU at dmax.

Show solution
kQ = 0.9882 + 1.486×10−3×66.4 − 2.140×10−5×66.4² = 0.9925
PTP = 294.2/295.2 × 101.33/101.8 = 0.9920
Pion = (1 − 2)/(14.70/14.67 − 2) = 1.0020   Ppol = 29.42/29.40 = 1.0007
M = 14.70 × 0.9920 × 1.0020 × 1.0007 × 1.000 = 14.62 nC
Dw(10) = 14.62×10−9 × 0.9925 × 4.50×107 = 0.6531 Gy per 100 MU
D(dmax) = 0.6531/0.664 = 0.9836 Gy per 100 MU

≈ 0.984 cGy/MU, 1.6% below 1.000

Sanity check: a Farmer chamber collects about 20 nC per Gy, and 0.65 Gy at 10 cm fits 14 to 15 nC. The result is outside the 1% annual tolerance, so physics adjusts the monitor chamber and repeats. Leaving out kQ gives 0.991; leaving out PTP also gives 0.991. Either error would hide most of a real 1.6% deficit.

EX 4

Lead foil and beam quality at 18 MV

For an 18 MV beam, the depth dose at 10 cm, 100 cm SSD, 10 × 10 cm, measured with a 1 mm lead foil 50 cm from the water surface, is %dd(10)Pb = 79.6%. Find %dd(10)x and kQ for an NE2571.

Show solution
%dd(10)x = [0.8905 + 0.00150 × 79.6] × 79.6 = 80.39%
kQ = 0.9882 + 1.486×10−3×80.39 − 2.140×10−5×80.39² = 0.9694

%dd(10)x ≈ 80.4%; kQ ≈ 0.969

Sanity check: %dd(10)x comes out slightly higher than the foil reading because the conversion removes the foil's own electron contribution. kQ near 0.97 is right for 18 MV. Skipping the conversion and using 79.6% gives kQ = 0.9709, a 0.15% error: small, but systematic in every patient treated on that beam.

EX 5

Electron calibration, 1999 formalism

An electron beam has I50 = 5.10 cm. A cylindrical chamber with kecal = 0.903 and ND,w = 4.50 × 107 Gy/C is placed at dref. Pgr = 0.997. The fully corrected reading for 100 MU is M = 24.40 nC. The clinical PDD at dref is 99.5%. Find R50, dref and the dose per MU at dmax. Use k′R50 = 0.9905 + 0.0710 e−R50/3.67.

Show solution
R50 = 1.029 × 5.10 − 0.06 = 5.19 cm;   dref = 0.6 × 5.19 − 0.1 = 3.01 cm
k′R50 = 0.9905 + 0.0710 e−5.188/3.67 = 1.0078
kQ = 0.997 × 1.0078 × 0.903 = 0.9073
Dw(dref) = 24.40×10−9 × 0.9073 × 4.50×107 = 0.9962 Gy
D(dmax) = 0.9962/0.995 = 1.0012 Gy per 100 MU

R50 ≈ 5.19 cm, dref ≈ 3.01 cm, ≈ 1.001 cGy/MU

Sanity check: R50 of 5.2 cm means a mean incident energy near 2.33 × 5.19 ≈ 12 MeV, and dref around 3 cm is right for a 12 MeV beam. Leaving out kecal gives 1.11 cGy/MU, an 11% error that no output check would accept. Using I50 as R50 puts dref at 2.96 cm.

EX 6

In vivo diode on warm skin

An in vivo diode calibrated at 22 °C reads 2.06 Gy on a patient whose expected entrance dose at dmax is 2.00 Gy. On skin the diode reaches 30 °C. Its sensitivity rises by 0.3% per °C. What is the temperature-corrected reading, and does it pass a 5% action level?

Show solution
Sensitivity at 30 °C = 1 + 0.003 × 8 = 1.024
Corrected = 2.06 / 1.024 = 2.012 Gy, +0.6% from expected

≈ 2.01 Gy; passes

Sanity check: a warmer diode is more sensitive and over-reads, so the correction must lower the reading. Multiplying instead gives 2.11 Gy, 5.5% high, a false alarm that would stop treatment for no reason. Diode readings also need SSD, field size and angle corrections in practice.

12 · Clinic layer

In the planning room

You will not run TG-51. You will sign plans whose every MU depends on it, read in vivo results, order film QA, and stand next to survey meters. These are the numbers and the questions that make those moments productive.

Ballpark numbers to carry in your head

QuantityRough numberWhat it tells you
Farmer chamber volume and signal0.6 cm³, ≈ 20 nC/GyTens of nC means about a gray
PTP per °C and per kPa≈ 0.34% and ≈ 1%A stormy day or a hot room moves the raw reading by percent
Pion limit≤ 1.05Above this, change chamber or voltage
Ppol expected0.996 to 1.004Outside it, the chamber or cable needs a look
kQ, Farmer, 6 MV and 18 MV≈ 0.99 and ≈ 0.97Below 1, falls with energy
TG-51 combined uncertainty≈ 1% (k = 1)The floor under every patient's dose
TG-142 output tolerances3% daily, 2% monthly, 1% annualWhen to hold treatment and call physics
ADCL recalibrationwithin 2 yearsTG-51 requirement for the chamber and electrometer

Choosing a detector for a clinical question

Is the TSET or chest wall skin dose right: OSLD or TLD on the skin, with no buildup cap, read against a calibration done under similar conditions. Is the pacemaker dose acceptable: OSLD or diode near the device, per the institutional policy for cardiac devices. Does an SRS output factor look odd: ask which detector measured it and whether the small-field correction was applied. Did the IMRT QA pass: the answer depends on the array or film, its resolution, and the gamma criteria (Module 11).

Say this to physics or dosimetryWhy it lands
"When was the last TG-51, and how does today's daily output compare?"Separates a machine drift from a one-off reading
"Was the in vivo diode corrected for SSD and temperature?"Explains most diode readings that are a few percent off
"What detector was used for these small-field output factors?"Volume averaging and over-response change small-field dose by several percent
"Are we on the 2024 electron addendum or the 1999 protocol?"Accounts for a step change of up to about 2% in electron output at recalibration
"Which meter did we survey the room with?"A GM counter can under-read near a high-activity source; an ion chamber meter is needed for dose rate
In clinic

Whole-breast tangents with a 5 mm bolus every other day: OSLDs on the skin under and away from bolus show you, in numbers, what the bolus does to surface dose. When the reading under bolus is near the prescription and the reading without it is around half, the detectors are telling you about the buildup region from Module 05, not about a problem with the plan.

13 · Rapid review

Board pearls and traps

Dw = M · kQ · ND,w

M = Pion PTP Pelec Ppol Pleak Mraw.

PTP = (273.2+T)/295.2 × 101.33/P

Above 1 when hot or at low pressure.

Pion ≥ 1, limit 1.05

Two-voltage method; rises with dose per pulse (FFF).

Ppol: average the two polarities

Largest for electrons and parallel-plate chambers.

Photon reference: 10 cm, 10 × 10, 100 SSD or SAD

Beam quality %dd(10)x; TRS-398 uses TPR20,10.

Lead foil, 1 mm

For %dd(10) ≈ 75% and above and for all FFF beams. Removed for the dose measurement.

kQ < 1, falls with energy

≈ 0.99 at 6 MV, ≈ 0.97 at 18 MV for a Farmer.

Electrons: R50 and dref = 0.6 R50 − 0.1

R50 = 1.029 I50 − 0.06 cm.

Effective point: 0.6 rcav photons, 0.5 rcav electrons

For depth curves; not applied at the photon reference point.

Primary standard: calorimetry

NIST to ADCL to clinic. Recalibrate within 2 years.

GM: sensitive, paralyzes

Ion chamber survey meter: accurate rate, flat energy response.

Neutrons: BF3 or He-3 in polyethylene

Moderate, capture, count charged products.

Common traps
  • Inverting PTP; hot or low-pressure days need a factor above 1.
  • Putting the low-voltage reading over the high-voltage reading in Pion and getting a value below 1.
  • Using %dd(10)x instead of the clinical %dd(10) to carry dose from 10 cm to dmax.
  • Leaving the lead foil in place for the dose measurement.
  • Calling a GM counter reading of zero next to a source proof that the source is safe.
  • Treating a capped in vivo diode reading as skin dose.
  • Using a Farmer chamber for a 1 cm SRS output factor.
14 · Test yourself

Question bank

Exam-style single best answer. Pick one to see the explanation.

0 / 0 answered
15 · Keep this

Formula card and sources

TG-51 dose
Dw = M · kQ · ND,w
M = Pion PTP Pelec Ppol Pleak Mraw
Air density
PTP = (273.2+T)/295.2 × 101.33/P
Recombination and polarity
Pion = (1 − VH/VL)/(MH/ML − VH/VL)
Ppol = |M+ − M−| / 2|M|
Photon beam quality
%dd(10)x = [0.8905 + 0.00150 %dd(10)Pb] %dd(10)Pb
kQ = A + B·10⁻³·x + C·10⁻⁵·x²
Electron beam quality
R50 = 1.029 I50 − 0.06 cm
dref = 0.6 R50 − 0.1 cm
1999: kQ = Pgr · k′R50 · kecal
Transfer to dmax
SSD: D(dmax) = D(10)/[%dd(10)/100]
Electrons: D(dmax) = D(dref)/PDD(dref)

Sources

  1. American Board of Radiology. Study Guide for Medical Physics for Radiation Oncology (2023 domain ranges; domain II.B, radiation measurement and calibration).
  2. Almond PR et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847-1870.
  3. McEwen M et al. Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon beams. Med Phys 2014;41:041501.
  4. Muir B et al. AAPM WGTG51 Report 385: Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy electron beams. Med Phys 2024;51(9):5840-5857.
  5. Klein EE et al. Task Group 142 report: Quality assurance of medical accelerators. Med Phys 2009;36:4197-4212.
  6. IAEA Technical Reports Series 398. Absorbed Dose Determination in External Beam Radiotherapy. Vienna, 2000, and its 2024 revision.
  7. IAEA TRS-483. Dosimetry of Small Static Fields Used in External Beam Radiotherapy. Vienna, 2017.
  8. Khan FM, Gibbons JP. The Physics of Radiation Therapy, 5th ed. Chapters on measurement of ionizing radiation and dosimetry instruments.
  9. Attix FH. Introduction to Radiological Physics and Radiation Dosimetry. Wiley, 1986. Chapters on ionization chambers and integrating dosimeters.
  10. US Nuclear Regulatory Commission. 10 CFR 35.61, Calibration of survey instruments; 10 CFR 35.630, Dosimetry equipment.

The worksheet applies the published photon formalism to invented readings. Clinical numbers come from your physicist's calibration, the chamber's ADCL report and your machine's commissioning data.