Module 09 · ABR domain III · Treatment planning

Electron Beams

Electrons treat a slab of tissue and then stop. That single property makes them the tool for skin cancers, scars, chest walls and superficial nodes, and it is also why their physics behaves so differently from photons: the beam scatters in air, bulges and constricts with depth, shortens in bone, lengthens in lung, and hates air gaps. This module teaches the rules of thumb the exam tests and the judgement you need when choosing an energy, a cutout and a bolus.

In this module
  1. 01The one idea
  2. 02Where the beam comes from
  3. 03The depth dose curve
  4. 04Field size, isodoses, cutouts
  5. 05Virtual source and air gaps
  6. 06Obliquity and surface shape
  7. 07Lung, bone and air
  8. 08Bolus and shielding
  9. 09Field matching
  10. 10Electron bench
  11. 11Worked examples
  12. 12In the planning room
  13. 13Board pearls and traps
  14. 14Question bank
  15. 15Formula card and sources
01 · Start here

The one idea

An electron beam is a charged particle beam with a finite range: it delivers a nearly flat dose to a depth set by its energy, then falls off steeply, and almost nothing reaches beyond. Every rule in this module follows from that range and from how easily electrons scatter.

Photons are uncharged. They cross tissue until they happen to interact, so a photon beam never stops; it only thins out exponentially (Module 03). Electrons are charged. They interact with every atom they pass, losing about 2 MeV for each centimetre of water, so a beam of a given energy runs out of energy at a predictable depth. That is the range.

Because electrons are light, every interaction also deflects them. They scatter in the air between the machine and the skin, they scatter sideways in tissue, and they scatter differently in bone, lung and air. Range explains the depth dose; scatter explains almost everything else: the shape of isodose lines, the loss of output in small fields, the sensitivity to air gaps and the hot and cold spots near heterogeneities.

Practical range Rp ≈ E (MeV)2 cm of water
About 2 MeV lost per cm of water. A 12 MeV beam stops at roughly 6 cm. The therapeutic depth (90 percent) is roughly a third to a quarter of the energy in MeV.
Pearl

Divide the energy by 4, 3 and 2: about R90, R80 and Rp in cm. For 12 MeV: 3, 4 and 6 cm. Most electron board questions are one of these three divisions.

02 · The source

Where the beam comes from

In electron mode the linac (Module 02) removes the x-ray target and the flattening filter from the beam path. The narrow pencil of electrons leaving the bending magnet is spread into a broad, flat field by scattering foils: a thin high-Z primary foil that widens the beam, and a shaped secondary foil that flattens it. The foils, the monitor chambers and the air also produce bremsstrahlung x-rays, which is the origin of the photon tail at the end of every electron depth dose curve.

Electrons scatter so much in air that the jaws, about 60 cm above the patient, cannot define a sharp field edge. An applicator (cone) carries the collimation down to about 5 cm above the skin. The field is shaped by a cutout, a Cerrobend or lead insert in the end of the applicator. The jaws are set automatically to a larger opening matched to each applicator, which keeps scatter from the applicator walls consistent.

Three energies you will see named

QuantityMeaningHow it is obtained
Nominal energyThe label on the console (6, 9, 12, 16, 20 MeV)A name, not a measurement
Ep,0, most probable energy at the surfaceThe peak of the energy spectrum at the phantom surfaceEp,0 = 0.22 + 1.98 Rp + 0.0025 Rp²
E0, mean energy at the surfaceAverage energy at the surfaceE0 ≈ 2.33 × R50
Ez, mean energy at depth zFalls nearly linearly with depth (Harder)Ez = E0 (1 − z / Rp)

A 12 MeV beam with R50 = 5.15 cm has E0 = 2.33 × 5.15 = 12.0 MeV. At 3 cm depth, with Rp = 6 cm, its mean energy is 12 × (1 − 3/6) = 6 MeV. The energy at depth matters whenever something sits inside the patient: an internal lead shield, a bone surface, a detector.

Board point

Beam quality for electron calibration is specified by R50, measured with I50 converted by R50 = 1.029 I50 − 0.06 cm, and the TG-51 reference depth is dref = 0.6 R50 − 0.1 cm (Module 06). E0 = 2.33 R50 is the bridge between the calibration number and the clinical energy.

03 · Depth

The depth dose curve

Read the curve in four parts, each with its own physics.

Surface. Electrons deposit dose from the moment they enter, so there is no real skin sparing. Surface dose is high, typically about 75 to 80 percent at 6 MeV and about 90 percent or more at 16 to 20 MeV. It rises with energy, the opposite of photons. Low-energy electrons scatter through larger angles near the surface, so the dose builds up more over the first centimetre (obliquely travelling electrons cross a thin layer with a longer path and deposit more); high-energy electrons travel straighter, so the surface already sits close to the maximum.

Build-up and plateau. The modest rise to dmax comes from electrons being scattered out of their straight paths, which increases fluence through each layer. dmax is broad and does not move deeper in proportion to energy; it plateaus at around 2.5 to 3 cm for higher energies and can even move shallower.

Fall-off. As electrons run out of energy, dose drops steeply. The gradient is sharpest at low energy and becomes more gradual at high energy, because range straggling and scatter spread the stopping points.

Bremsstrahlung tail. Beyond Rp, a few percent of dose remains from x-rays produced in the foils, collimation and the patient. It grows with energy because radiative losses scale with energy.

R90 ≈ E/4    R80 ≈ E/3    R50 ≈ E/2.33    Rp ≈ E/2   (cm, E in MeV)
Rules of thumb for a broad beam in water. Measured beams usually penetrate a little farther than E/4 and E/3 (R90 nearer E/3.2, R80 nearer E/2.8), so the board rules err on the conservative side. Your commissioned depth doses govern.
NominalSurfacedmaxR90R80RpX-ray tail
6 MeV≈ 75 to 80%≈ 1.3 cm≈ 1.8 cm≈ 2.0 cm≈ 3.0 cm≈ 0.5 to 1%
9 MeV≈ 80%≈ 2.0 cm≈ 2.7 cm≈ 3.0 cm≈ 4.4 cm≈ 1%
12 MeV≈ 85%≈ 2.7 cm≈ 3.8 cm≈ 4.2 cm≈ 6.0 cm≈ 1 to 2%
16 MeV≈ 90%≈ 3.0 cm≈ 5.1 cm≈ 5.6 cm≈ 8.0 cm≈ 2 to 3%
20 MeV≈ 90 to 95%≈ 2.5 cm≈ 6.0 cm≈ 6.8 cm≈ 10 cm≈ 4 to 5%

Typical values for a 10 x 10 applicator at 100 cm SSD, approximate. Linac models differ by several millimetres; your machine's commissioning data take precedence.

Increase thisSurface dosePenetration (R80, Rp)Distal gradientX-ray tail
EnergyRisesDeeperShallower slopeRises
Field size, below lateral equilibriumLittle changeDeeper, until equilibriumLittle changeSlight rise
SSD (air gap)Little changeNearly unchangedLittle changeLittle change
Beam obliquityRisesShallowerLess steepLittle change
Board point

Electron surface dose rises with energy; photon surface dose falls with energy. Electron PDD is nearly independent of SSD over the clinical range, so the SSD effect for electrons shows up in output, not in the shape of the curve.

In clinic

Choose energy so the 90 percent line (some clinics accept 80 percent) reaches the deep edge of the target, then check where Rp lands. Over a chest wall that means lung; over the posterior neck, cord; over the scalp, brain. Rp is the depth below which the structure receives only the tail.

04 · Width

Field size, isodose shape and cutouts

Each electron entering the patient spreads sideways as it goes deeper. A point on the central axis of a large field receives as many electrons scattered in from the sides as it loses to scatter out. That balance is lateral scatter equilibrium. When the field edge is closer to the axis than the electrons can travel sideways, more leave than return, and the central axis dose falls.

Equilibrium needs the field edge to be farther from the point than the lateral spread of the electrons. Khan quotes the radius needed as Req ≈ 0.88 √Ep,0 cm, about 3 cm at 12 MeV. A practical rule follows: once the smallest field dimension exceeds about the practical range Rp, depth dose no longer depends on field size.

As the cutout shrinks below equilibriumWhat happensWhy
Output (dose per MU at dmax)FallsFewer electrons scatter in from the sides
dmaxMoves shallowerLateral loss grows with depth, so the peak forms earlier
PDD at depth, R90, R80Fall, shallowerThe deep dose depends most on in-scatter
Relative surface doseRisesThe peak falls toward the surface value

Isodose shape: bulge and constriction

Lateral scatter pushes low-value isodose lines (10 to 20 percent) outward beyond the geometric field edge, especially at higher energies. The high-value lines (80 to 90 percent) are pulled inward with depth, because points near the edge lose electrons and are not compensated. The 90 percent line is therefore narrower at depth than the field at the surface.

Electron isodose shape (schematic) field edgefield edge 90% 20% low lines bulge out high lines pull in 90% width at depth < field width

The 90 percent line constricts with depth and the low-dose lines bulge beyond the field edge. Margins drawn on the skin must be wider than the target at depth.

In clinic

Draw the cutout about 1 cm beyond the target on each side for typical energies, more for deeper targets and higher energies, so the target at depth sits inside the constricted 90 percent line. For a skin cancer with a clinical margin already drawn, the cutout adds this dosimetric margin on top.

Cutout output factors

Output for an electron field is measured for each applicator and for each cutout size, relative to the reference applicator and cutout. For a rectangular cutout that has not been measured, the square root rule estimates its output from two square fields.

OF(X × Y) ≈ √( OF(X × X) × OF(Y × Y) )
Example: OF(4 × 4) = 0.92 and OF(10 × 10) = 1.000 give OF(4 × 10) ≈ √0.92 = 0.959. Irregular and very small cutouts should be measured.
Trap

Using the photon equivalent square (4A/P) for an electron cutout. Electron output does not follow Sterling's rule closely; the square root method or a direct measurement is the electron answer.

05 · Distance

Virtual source and air gaps

A photon beam comes from a small spot on the target, so inverse square works from the nominal source position. An electron beam has no such spot. The scattering foils, the monitor chambers, the air and the applicator spread the electrons so that, seen from the patient, they appear to come from a point closer than the nominal source. Two different "sources" are used, for two different jobs.

ConceptDefined byUsed for
Virtual sourceBack-projecting the 50 percent field edges measured in air at several distancesField size and penumbra geometry at extended distance
Effective SSDThe distance that makes inverse square fit the measured output versus air gapOutput (MU) correction at extended SSD

The effective SSD is measured by placing a chamber at dmax in phantom and recording the reading I0 with no gap and Ig with an air gap g between the applicator end and the surface. If inverse square holds from an effective source, the square root of the reading ratio is linear in the gap.

√( I0 / Ig ) = 1 + gSSDeff + dmax
Output at extended SSD = ( SSDeff + dmaxSSDeff + g + dmax )²
Plot √(I0/Ig) against g: the slope is 1/(SSDeff + dmax). The effective SSD is shorter than the nominal 100 cm and is lower for low energies and small cutouts; it is measured for each energy and applicator.

Because the effective source is closer, output falls faster with an air gap than the nominal inverse square predicts. Using 100 cm instead of an effective SSD near 85 cm underestimates the loss and under-doses the patient, by about 1.5 percent for a 5 cm gap at 12 MeV and more for larger gaps.

An air gap also widens the penumbra, because electrons keep scattering sideways in the extra air. Depth dose shape changes little.

Board point

Effective SSD is less than nominal SSD. Output correction at an extended distance uses the effective SSD in the inverse square factor. Penumbra widens with air gap. PDD is nearly independent of SSD.

In clinic

Shoulders, the nose and the curved scalp often force a standoff. Ask what gap the setup will have and whether the planning or hand calc uses the measured effective SSD for that applicator. Beyond a few centimetres of gap, consider skin collimation to keep the edge sharp.

06 · The real surface

Obliquity and surface shape

When the beam strikes the skin obliquely, electrons travel along slanted paths. The slant packs more electron track into each shallow layer and also lets electrons scattered from neighbouring regions add dose near the surface. Dose near the surface and at dmax rises, dmax moves shallower, and the dose at depth falls because the path to a given perpendicular depth is longer. The penetration measured perpendicular to the skin shrinks.

Surface shape matters on a smaller scale. Electrons scatter preferentially from the thicker side of a step toward the thinner side. Below a sharp ridge or the edge of a depression (the nasal ala, the ear, a surgical defect), this produces adjacent hot and cold spots that a pencil beam calculation may not show.

Surface featureEffect on doseFix
Oblique incidence, curved chest wall or scalpHotter shallow, shallower penetrationSplit fields, change gantry angle, bolus to flatten the surface
Sharp irregularity (nose, ear, surgical step)Local hot and cold spotsBolus to fill the irregularity
Surface curving away from the applicatorLarger gap at the field edge, lower output, wider penumbraShorter standoff, skin collimation
Trap

Assuming an oblique electron beam penetrates deeper because the slant path through tissue is longer. It does the opposite: the electron range is spent along the slant, so the perpendicular depth reached is shallower.

07 · Heterogeneities

Lung, bone and air

Electrons lose energy in proportion to the electron density they cross. A centimetre of lung removes far less energy than a centimetre of water, so the beam travels farther; a centimetre of compact bone removes more, so the beam stops sooner. The simplest correction is the coefficient of equivalent thickness (CET): the thickness of water that has the same effect on the beam as 1 cm of the material.

deff = d − t (1 − CET)
d is the physical depth of the point, t the thickness of the inhomogeneity above it. Read the depth dose at deff. CET is below 1 in lung (the point behaves as if shallower, so dose beyond lung is higher and the beam reaches deeper) and above 1 in bone.
MaterialApproximate CETConsequence
Lung≈ 0.5 (varies with density and depth)Range extends about twice as far per cm crossed
Spongy bone≈ 1.1Small shortening
Compact bone≈ 1.65Range shortens noticeably behind cortical bone
Air cavity≈ 0The beam crosses it almost for free; dose beyond rises

Values after Khan and Gibbons, approximate. Lung CET is not simply the physical density: it reflects electron transport as well as stopping power.

CET corrects depth on the central axis of a broad inhomogeneity. At the edges of a small inhomogeneity, scatter creates its own pattern: more electrons scatter out of the denser medium than into it, so a hot spot forms on the low-density side of the interface and a cold spot on the high-density side. Behind the edge of a bone or an air cavity, these hot and cold streaks can be large. Monte Carlo calculation (TG-105) handles them; a pencil beam algorithm handles them poorly.

Internal shields and backscatter

When lead sits inside the patient (an eye shield under the lid, a lip or buccal mucosa shield), electrons striking it backscatter into the tissue in front of it. The dose upstream of the lead rises substantially, by tens of percent at the energies used for these sites, and more at lower electron energy at the interface. The usual fix is a coating of low-Z material, such as wax, dental acrylic or aluminium, on the beam side of the lead, a few millimetres thick, to absorb the backscattered electrons.

Board point

Electron backscatter from lead increases dose on the entrance side of an internal shield and is larger at lower energy at the interface. A low-Z coating absorbs it. The mean energy at the shield depth comes from Ez = E0(1 − z/Rp).

In clinic

Postmastectomy chest wall: the chest wall is thin and the lung immediately deep to it extends the range. A 12 MeV beam with an 80 percent depth of 4 cm in water, crossing 1.5 cm of chest wall, reaches 1.5 + 2.5/0.5 = 6.5 cm physical depth with its 80 percent line. Pick the lowest energy that covers the chest wall, and look at the lung dose on the plan.

08 · Pulling and stopping the beam

Bolus and shielding

Bolus is tissue-equivalent material on the skin. For electrons it does two jobs. It raises surface dose, by moving the skin into the high part of the curve. And it pulls the whole distribution toward the surface by its thickness, which is how you treat a superficial lesion with an energy whose range would otherwise reach a critical structure. A variable-thickness bolus shaped to the distal target surface (bolus electron conformal therapy) uses the second job to sculpt the 90 percent line.

Add bolus of thickness tEffect
Surface doseRises toward the dmax level
R90, R80, Rp in the patientAll move shallower by t
Dose to the structure beneathFalls

How much lead stops an electron beam

Electrons have a finite range in lead as in water, but lead also converts some of their energy into bremsstrahlung, so a shield can never reach zero. The rule for the minimum lead thickness that brings transmission to about 5 percent is simple.

Lead thickness (mm) ≈ E (MeV)2
Use the mean energy at the shield. Add about 1 mm for safety. Cerrobend needs about 20 percent more thickness than lead. Too thin a shield can raise dose under it above the open-field dose, because it scatters and adds bremsstrahlung without stopping the beam.
Pearl

"Half the MeV in millimetres of lead." A 9 MeV eyelid field needs about 4.5 mm; a 6 MeV field about 3 mm. Internal eye shields are often thin, so they are used only with low energies.

Trap

Using a bolus to raise skin dose for an electron field and forgetting it also pulls R90 shallower by the same amount. A 1 cm bolus on a lesion 2 cm thick needs coverage to 3 cm of water-equivalent depth from the bolus surface.

09 · Junctions

Field matching

Photon junctions are a matter of divergence (Module 08). Electron junctions add scatter. Because low-value isodose lines bulge beyond the field edge, two electron fields abutted at the skin overlap at depth and produce a hot spot below the junction; separating them on the skin produces a cold strip at the surface. There is no perfect gap, so electron junctions are usually abutted with a planned shift (feathering) every few fractions, or aligned so the beam edges are parallel at the junction.

When an electron field abuts a photon field at the skin, the asymmetry is the board point. Electrons scatter out of the electron field into the photon field, where the photon dose is already full at depth. The result is a hot spot on the photon side and a cold spot on the electron side of the junction.

JunctionWhat appearsClinical example
Electron to electron, abutted at skinHot spot at depth under the junctionWide chest wall covered by two electron fields
Electron to electron, gappedCold strip near the surfaceSame, with a skin gap
Electron to photon, abutted at skinHot on the photon side, cold on the electron sidePosterior neck electron strips beside photon fields in older 3D head and neck plans; IMN electron field beside breast tangents
In clinic

Posterior neck electrons (often 9 to 12 MeV) were classically used to treat nodes behind the cord after the photon fields were pulled off the cord. The cord depth decides the energy: Rp must stop short of it. The junction hot spot falls on the photon side, close to the cord, so the match line and the cord dose are reviewed on every plan.

10 · Run it

Electron bench

Pick an energy, then add bolus, a slab of lung or bone, or an air gap, one at a time. Watch which part of the curve moves and say why before you read the numbers. The dashed curve is the same energy in plain water at 100 cm SSD.

Electron depth dose bench

Teaching model, not commissioning data
Nominal energy

Depths are physical depths below the skin. The slab correction uses CET on the central axis only; it does not show the edge hot and cold spots.

Your setupSame energy, water, no bolusSlab position
Skin dose0%
dmax0
R900
R800
R500
Rp0
X-ray tail0
E0 = 2.33 R50 (water)0
Effective SSD (model)0
Output vs 100 cm, effective SSD0
Output vs 100 cm, nominal ISF0
11 · Practice the method

Worked examples

Work each one on paper first. Factors are given as they are on the exam.

EX 1

Choose the energy

A recurrent nodule on the chest wall extends to 3.0 cm below the skin. You want the 90 percent line at the deep edge. Available energies: 6, 9, 12, 16 MeV. Using the board rules, which energy, and where does the practical range fall?

Show solution
R90 ≈ E/4 → E ≥ 4 × 3.0 = 12 MeV;   Rp ≈ 12/2 = 6 cm

12 MeV; Rp ≈ 6 cm, in water-equivalent depth

Sanity check: 9 MeV gives R90 ≈ 2.25 cm by the rule (about 2.7 cm on real beams), short of 3 cm. The common error is using E/3, which gives 9 MeV and covers the deep edge only at the 80 percent level. Then check what lies at 3 to 6 cm: on a chest wall, lung, where the range is extended further (EX 5).

EX 2

Electron MU

12 MeV, 10 x 10 applicator with a cutout whose output factor is 0.97, 100 cm SSD. Prescribe 200 cGy to the 90 percent isodose line. The machine gives 1 cGy/MU at dmax for the reference applicator at 100 cm SSD.

Show solution

Electron MU are calculated at dmax: the dose at dmax is 200 ÷ 0.90, and each MU gives 1 × 0.97 cGy there.

MU = 200 ÷ (1 × 0.97 × 0.90) = 229.1 MU

≈ 229 MU

Sanity check: prescribing to the 90 percent line always raises MU by 1/0.9 = 1.11. Leaving the isodose level out gives 206 MU, an 11 percent under-dose. Putting 0.90 in the numerator gives 186 MU.

EX 3

Extended SSD with effective SSD

The field in EX 2 must be treated at 105 cm SSD because of the shoulder. The effective SSD for this energy and applicator is 85 cm; dmax = 2.7 cm. Find the MU, and the error if nominal inverse square (100 cm) were used.

Show solution
Gap factor = [(85 + 2.7) ÷ (85 + 5 + 2.7)]² = (87.7 ÷ 92.7)² = 0.8950
MU = 229.1 ÷ 0.8950 = 256.0 MU
Nominal: [(100 + 2.7) ÷ (105 + 2.7)]² = 0.9093 → 251.9 MU

≈ 256 MU; nominal ISF would under-dose by about 1.6%

Sanity check: a 5 cm gap costs about 10 percent of output with the effective SSD, more than the 9 percent the nominal source predicts. The effective source is closer, so the same gap is a larger fraction of the distance.

EX 4

Measure the effective SSD

A chamber at dmax = 2.0 cm reads 100.0 with the phantom surface at the applicator end (no gap) and 80.4 with a 10 cm air gap. Find the effective SSD.

Show solution
√(100.0 ÷ 80.4) = 1.1153 → slope = 0.1153 ÷ 10 = 0.01152 per cm
SSDeff + dmax = 1 ÷ 0.01152 = 86.8 cm → SSDeff = 84.8 cm

SSDeff ≈ 85 cm

Sanity check: with nominal 100 cm, a 10 cm gap would leave (102/112)² = 0.83 of output, i.e. a reading near 83. The measured 80.4 is lower, so the effective source must be closer than 100 cm. Forgetting to subtract dmax gives 86.8 cm.

EX 5

Range in lung

A 12 MeV chest wall field crosses 1.5 cm of soft tissue, then lung with CET 0.5. Using R80 ≈ E/3, at what physical depth does the 80 percent line lie?

Show solution
R80 ≈ 12/3 = 4.0 cm water-equivalent
Water-equivalent depth left after the chest wall: 4.0 − 1.5 = 2.5 cm
Physical depth in lung: 2.5 ÷ 0.5 = 5.0 cm → total 1.5 + 5.0 = 6.5 cm

≈ 6.5 cm below the skin

Sanity check: lung lets the beam travel about twice as far per centimetre, so the 80 percent line must sit deeper than 4 cm, not shallower. Multiplying by the CET instead of dividing gives 2.75 cm, which wrongly suggests lung shortens the range.

EX 6

Size an eyelid shield

A 9 MeV field treats a lower eyelid lesion. What minimum lead thickness blocks the beam over the lens, and how thick would the equivalent Cerrobend cutout be?

Show solution
Lead: 9 ÷ 2 = 4.5 mm (about 5.5 mm with a 1 mm margin)
Cerrobend: 1.2 × 4.5 = 5.4 mm minimum

≈ 4.5 mm lead minimum; ≈ 5.4 mm Cerrobend

Sanity check: millimetres of lead, not centimetres; electrons stop quickly in high-Z material. For an internal eye shield, the energy at the shield depth is lower than nominal, but backscatter from the shield raises lid dose, so the shield is coated on the beam side.

12 · Clinic layer

In the planning room

Ballpark numbers to carry in your head

SituationRough numberWhat it tells you
Therapeutic depth (R90)E/4 to E/3.2 cm6 MeV treats about 1.5 to 2 cm; 12 MeV about 3 to 4 cm
Practical rangeE/2 cmWhere the structure underneath stops being dosed
Surface dose without bolus≈ 75 to 95%Low energies spare skin a little; add bolus for full skin dose
Field margin beyond target at the skin≈ 1 cm or moreThe 90 percent line constricts with depth
Lead for shieldingE/2 mmPlus about 1 mm; Cerrobend about 20 percent thicker
Air gap output loss≈ 2% per cm near 100 cmLarger with the effective SSD; use the measured value
Lung beyond the chest wallrange ≈ doubled per cmChoose the lowest energy that covers the chest wall
Say this to physics or dosimetryWhy it lands
"Where does R90 fall relative to the deep edge, and where does Rp fall relative to the lung or cord?"Coverage and sparing are both set by the energy
"Is the cutout small enough to lose lateral equilibrium? Was its output measured?"Small cutouts lose output and depth; the planning system may not model them well
"What is the standoff at setup, and does the calc use the effective SSD for this applicator?"Nominal inverse square under-doses at extended SSD
"Do we need bolus for skin dose, and has the energy been raised to account for it?"Bolus pulls R90 shallower by its thickness
"Is there a nasal cavity, sinus or bone edge under the field? Has it been calculated with a heterogeneity-aware algorithm?"Edge hot and cold spots are invisible to simple calcs
"Where is the electron to photon match, and where is the hot spot?"It lies on the photon side; know what it overlaps
In clinic

For a basal cell carcinoma on the nose treated with electrons, the usual recipe is a low energy (6 to 9 MeV), bolus to fill irregularities and bring skin to full dose, a cutout with a dosimetric margin, a nasal or intraoral shield coated on the beam side when needed, and the shortest standoff the anatomy allows. Orthovoltage x-rays are an alternative for very superficial lesions and avoid the electron problems of small fields and air gaps.

13 · Rapid review

Board pearls and traps

E/4, E/3, E/2

R90, R80, Rp in cm. R50 ≈ E/2.33.

About 2 MeV per cm

Electron energy loss in water; Ez = E0(1 − z/Rp).

E0 = 2.33 R50

Ep,0 = 0.22 + 1.98 Rp + 0.0025 Rp².

Surface dose rises with energy

Opposite to photons. About 75 to 80% at 6 MeV, 90% or more at 16 to 20 MeV.

X-ray tail rises with energy

About 0.5 to 1% at 6 to 9 MeV, up to about 4 to 5% at 20 MeV. Source: foils, collimation, patient.

Small cutouts

Output falls, dmax shallower, PDD at depth falls. Equilibrium once the field exceeds about Rp.

Isodoses

90% constricts with depth, 10 to 20% bulges out. Add about 1 cm margin at the skin.

Effective SSD < nominal

Use it for output at extended SSD. Penumbra widens with gap; PDD barely changes.

Obliquity

Surface and dmax dose up, dmax shallower, penetration down.

CET

Lung about 0.5, compact bone about 1.65. deff = d − t(1 − CET).

Lead: E/2 mm

Plus 1 mm. Cerrobend 1.2 times lead. Coat internal shields against backscatter.

Electron to photon junction

Hot on the photon side, cold on the electron side.

Square root rule

OF(X × Y) ≈ √(OF(X × X) · OF(Y × Y)).

TG-51 electrons

Quality R50; dref = 0.6 R50 − 0.1 cm (Module 06).

Common traps
  • Using E/3 when the question asks for the 90 percent depth. E/3 is R80.
  • Correcting extended SSD output with the nominal 100 cm source instead of the effective SSD.
  • Saying electron surface dose falls with energy, by analogy with photons.
  • Multiplying by the lung CET instead of dividing when finding the physical depth reached in lung.
  • Forgetting that bolus moves R90 shallower in the patient by its thickness.
  • Placing the electron to photon hot spot on the electron side.
  • Using Sterling's 4A/P equivalent square for an electron cutout.
  • Expecting an oblique electron beam to penetrate deeper.
14 · Test yourself

Question bank

Exam-style single best answer. Pick one to see the explanation.

0 / 0 answered
15 · Keep this

Formula card and sources

Range rules (cm, E in MeV)
R90 ≈ E/4 · R80 ≈ E/3
R50 ≈ E/2.33 · Rp ≈ E/2
Energies
E0 = 2.33 · R50
Ep,0 = 0.22 + 1.98 Rp + 0.0025 Rp²
Ez = E0 (1 − z/Rp)
MU
MU = D ÷ [K · OF · (IDL/100) · gap factor]
Effective SSD
√(I0/Ig) = 1 + g / (SSDeff + dmax)
gap factor = [(SSDeff+dm)/(SSDeff+g+dm)]²
Heterogeneity and cutouts
deff = d − t(1 − CET)
OF(X×Y) ≈ √(OF(X×X) · OF(Y×Y))
Shielding
Pb (mm) ≈ E/2, +1 mm
Cerrobend ≈ 1.2 × Pb

Sources

  1. American Board of Radiology. Radiation Oncology Qualifying Exam: Study Guide for Medical Physics for Radiation Oncology (2023 domain ranges).
  2. Khan FM, Gibbons JP. The Physics of Radiation Therapy, 5th ed. Chapter on electron beam therapy.
  3. Khan FM et al. Clinical electron-beam dosimetry: report of AAPM Radiation Therapy Committee Task Group 25. Med Phys 1991;18:73-109.
  4. Gerbi BJ et al. Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25 (TG-70). Med Phys 2009;36:3239-3279.
  5. Almond PR et al. AAPM TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847; Muir B et al. WGTG51 Report 385, electron addendum. Med Phys 2024;51:5840.
  6. Hogstrom KR, Almond PR. Review of electron beam therapy physics. Phys Med Biol 2006;51:R455-R489.
  7. Ford E. Primer on Radiation Oncology Physics. CRC Press, 2020.

The bench uses a simplified depth dose model fitted to typical published 6 to 20 MeV behaviour and assumed effective SSDs. It teaches direction and magnitude. Clinical numbers come from your machine's commissioning data.